Feasibility of a customizable training environment for neurointerventional skills assessment

PLoS One. 2020 Sep 17;15(9):e0238952. doi: 10.1371/journal.pone.0238952. eCollection 2020.

Abstract

Objective: To meet increasing demands to train neuroendovascular techniques, we developed a dedicated simulator applying individualized three-dimensional intracranial aneurysm models ('HANNES'; Hamburg Anatomic Neurointerventional Endovascular Simulator). We hypothesized that HANNES provides a realistic and reproducible training environment to practice coil embolization and to exemplify disparities between neurointerventionalists, thus objectively benchmarking operators at different levels of experience.

Methods: Six physicians with different degrees of neurointerventional procedural experience were recruited into a standardized training protocol comprising catheterization of two internal carotid artery (ICA) aneurysms and one basilar tip aneurysm, followed by introduction of one framing coil into each aneurysm and finally complete coil embolization of one determined ICA aneurysm. The level of difficulty increased with every aneurysm. Fluoroscopy was recorded and assessed for procedural characteristics and adverse events.

Results: Physicians were divided into inexperienced and experienced operators, depending on their experience with microcatheter handling. Mean overall catheterization times increased with difficulty of the aneurysm model. Inexperienced operators showed longer catheterization times (median; IQR: 47; 30-84s) than experienced operators (21; 13-58s, p = 0.011) and became significantly faster during the course of the attempts (rho = -0.493, p = 0.009) than the experienced physicians (rho = -0.318, p = 0.106). Number of dangerous maneuvers throughout all attempts was significantly higher for inexperienced operators (median; IQR: 1.0; 0.0-1.5) as compared to experienced operators (0.0; 0.0-1.0, p = 0.014).

Conclusion: HANNES represents a modular neurointerventional training environment for practicing aneurysm coil embolization in vitro. Objective procedural metrics correlate with operator experience, suggesting that the system could be useful for assessing operator proficiency.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Blood Vessel Prosthesis
  • Catheterization / methods
  • Cerebral Angiography / methods
  • Computer Simulation
  • Education, Medical / methods*
  • Embolization, Therapeutic / methods*
  • Feasibility Studies
  • Female
  • Humans
  • Intracranial Aneurysm / therapy
  • Male
  • Middle Aged
  • Models, Biological
  • Simulation Training / methods*

Grants and funding

This work was supported by the German Ministry for Education and Research (BMBF Grant 031L0068B). The sponsor encourages publication but had no role in the study design, data collection and analysis, decision to publish, or preparation of the manuscript.