We evaluated the ability of different fluorescent indicators by various analytical instruments, including a laser scanning confocal microscope (LSCM), fluorescence plate reader, and flow cytometer (FCM), to measure the mitochondrial membrane potential (ΔΨm) of cardiac H9c2 cells during oxidative stress-induced mitochondrial injury. The mitochondrial oxygen consumption rate and a transmission electron microscope were used to detect changes in mitochondrial functions and morphology, respectively. Cardiac H9c2 cells were exposed to H2O2 (500, 750, 1000, and 1250 μM) to induce mitochondrial oxidative stress injury, and fluorescent indicators including tetramethyl rhodamine ethyl ester (TMRE), 5,5',6,6'-tetrachloro-1,1',3,3'-tetraethylbenzimidazolocarbocyanine iodide (JC-1), and rhodamine 123 (R123) were used to detect changes in ΔΨm using an LSCM, fluorescence plate reader, and FCM. The decrease in ΔΨm caused by H2O2 was determined by endpoint and dynamic analyses after staining with JC-1 or TMRE. With the R123 probe, the LSCM could only detect the change in ΔΨm caused by 1000 μM H2O2. Moreover, R123 was less effective than JC-1 and TMRE for measurement of ΔΨm by the LSCM. Our data indicated that an LSCM is the most suitable instrument to detect dynamic changes in ΔΨm, whereas all three instruments can detect ΔΨm at the endpoint.