In-situ regeneration of P-doped optical fiber dosimeter

Opt Lett. 2020 Sep 15;45(18):5201-5204. doi: 10.1364/OL.402382.

Abstract

We demonstrate the feasibility of resetting and reusing dosimeters exploiting the measurement of the infrared radiation-induced attenuation (IR-RIA) in phosphosilicate optical fibers (OFs) to provide point or distributed dose measurements in radiation environments. To bleach the room temperature stable IR-RIA, we used the photobleaching (PB) phenomenon. The PB efficiency was evaluated for different wavelengths in the [400-1100] nm range. The best identified PB resetting condition consists in using a continuous-wave Argon-ion laser at 514 nm. This treatment successfully bleached ∼97% of the IR-RIA at 1550 nm of a 30 m-long P-doped single mode optical fiber X-ray irradiated at a dose of 100 Gy. Successive re-irradiations of the same OF sample, regenerated after each run, confirm that the dosimeter keeps the same calibration curve during the whole process.