Comparison between 2D and 3D MEDIC for human cervical spinal cord MRI at 3T

J Med Radiat Sci. 2021 Mar;68(1):4-12. doi: 10.1002/jmrs.433. Epub 2020 Sep 15.

Abstract

Introduction: High-resolution magnetic resonance imaging (MRI) of the cervical spinal cord is important to provide accurate diagnosis and pathological assessment of injuries. MEDIC (Multiple Echo Data Image Combination) sequences have been used in clinical MRI; however, a comparison of the performance of 2D and 3D MEDIC for cervical spinal cord imaging has not been reported. The aim of this study is to compare axial 2D and 3D MEDIC for the visualisation of the grey matter (GM) and white matter (WM) of the human cervical spinal cord.

Methods: Eight healthy participants were scanned using Siemens Prismafit 3T MRI. T2*-weighted gradient spoiled 2D and 3D MEDIC sequences were acquired at 0.4 × 0.4 × 3.0 and 0.3 × 0.3 × 3.0 mm resolutions, with the acquisition times of 6 and 7 min, respectively. Quantitative analyses of the images were made based on the image signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR) and non-uniformity (NU). Two independent radiologists (CS and FN), each provided Likert scoring assessments of anatomical visibility of the GM and WM structures and image clarity for all samples.

Results: Quantitative evaluation showed that 3D MEDIC provided higher SNR, higher CNR and lower NU than 2D MEDIC. However, 2D MEDIC provided better anatomical visibility for the GM, WM and CSF, and higher image clarity (lower artefacts) compared to 3D MEDIC.

Conclusions: 2D MEDIC provides better information for depicting the internal structures of the cervical spinal cord compared to 3D MEDIC.

Keywords: 3 Tesla; MEDIC; cervical spinal cord; grey matter; white matter.

Publication types

  • Comparative Study

MeSH terms

  • Adult
  • Artifacts
  • Cervical Cord / diagnostic imaging*
  • Female
  • Humans
  • Image Processing, Computer-Assisted
  • Magnetic Resonance Imaging / methods*
  • Male
  • Middle Aged
  • Young Adult