Imaging of Stroke in Rodents Using a Clinical Scanner and Inductively Coupled Specially Designed Receiver Coils

Ann Biomed Eng. 2021 Feb;49(2):746-756. doi: 10.1007/s10439-020-02610-0. Epub 2020 Sep 11.

Abstract

Imaging of small laboratory animals in clinical MRI scanners is feasible but challenging. Compared with dedicated preclinical systems, clinical scanners have relatively low B0 field (1.5-3.0 T) and gradient strength (40-60 mT/m). This work explored the use of wireless inductively coupled coils (ICCs) combined with appropriate pulse sequence parameters to overcome these two drawbacks, with a special emphasis on the optimization of the coil passive detuning circuit for this application. A Bengal rose photothrombotic stroke model was used to induce cortical infarction in rats and mice. Animals were imaged in a 3T scanner using T2 and T1-weighted sequences. In all animals, the ICCs allowed acquisition of high-quality images of the infarcted brain at acute and chronic stages. Images obtained with the ICCs showed a substantial increase in SNR compared to clinical coils (by factors of 6 in the rat brain and 16-17 in the mouse brain), and the absence of wires made the animal preparation workflow straightforward.

Keywords: Clinical scanner; Inductively coupled coil; Preclinical imaging; Stroke; Wireless coil.