Development of high-resolution YAP(Ce) x-ray camera for the imaging of astatine-211(At-211) in small animals

Med Phys. 2020 Nov;47(11):5739-5748. doi: 10.1002/mp.14455. Epub 2020 Sep 22.

Abstract

Purpose: Astatine-211 (At-211) is a promising alpha emitter for radionuclide therapy. High-resolution in vivo imaging of At-211 in small animals is needed for the development of At-211 radiopharmaceuticals. For this purpose, we developed a low-energy x-ray camera using a thin YAlO3 :Ce (YAP(Ce)) plate to image the low-energy x rays (73-87 keV) from the daughter radionuclide of At-211 (Po-211).

Method: We optically coupled a 38 × 38 × 1-mm YAP (Ce) plate to a position-sensitive photomultiplier (PSPMT) to develop an imaging detector. A pinhole or a parallel hole collimator was attached to the imaging detector, and the performance was measured for 60-keV gamma photons. With the developed x-ray camera, we carried out imaging of a mouse that had been administered At-211-NaAt.

Results: The intrinsic spatial resolution of the YAP (Ce) x-ray camera was approximately 1.2 mm FWHM, and the energy resolution was 22% FWHM. With a 5-mm-thick parallel hole collimator, the spatial resolution was 3.8 mm FWHM with a sensitivity of 8 × 10-4 at 10 mm, which is a typical distance from the surface of the collimator to a subject in mouse imaging. Using a 1-mm diameter pinhole collimator, the spatial resolution was 1.8 mm FWHM with a sensitivity of 3.5 × 10-4 at 10 mm from the collimator. In the mouse images measured by the developed x-ray camera, we could clearly observe that the At-211 accumulated in the thyroid gland and the stomach of the mouse.

Conclusion: We concluded that the YAP (Ce) x-ray camera is useful for in vivo imaging of At-211.

Keywords: 211At; YAP(Ce); characteristic x rays; imaging; in vivo imaging.

MeSH terms

  • Animals
  • Astatine
  • Gamma Cameras*
  • Mice
  • Phantoms, Imaging
  • Radiography
  • X-Rays

Substances

  • Astatine-211
  • Astatine