Augmented reality with HoloLens in parotid surgery: how to assess and to improve accuracy

Eur Arch Otorhinolaryngol. 2021 Jul;278(7):2473-2483. doi: 10.1007/s00405-020-06351-7. Epub 2020 Sep 10.

Abstract

Purpose: Augmented reality improves planning and execution of surgical procedures. The aim of this study was to evaluate the feasibility of a 3D augmented reality hologram in live parotic surgery. Another goal was to develop an accuracy measuring instrument and to determine the accuracy of the system.

Methods: We created a software to build and manually align 2D and 3D augmented reality models generated from MRI data onto the patient during surgery using the HoloLens® 1 (Microsoft Corporation, Redmond, USA). To assess the accuracy of the system, we developed a specific measuring tool applying a standard electromagnetic navigation device (Fiagon GmbH, Hennigsdorf, Germany).

Results: The accuracy of our system was measured during real surgical procedures. Training of the experimenters and the use of fiducial markers significantly reduced the accuracy of holographic system (p = 0.0166 and p = 0.0132). Precision of the developed measuring system was very high with a mean error of the basic system of 1.3 mm. Feedback evaluation demonstrated 86% of participants agreed or strongly agreed that the HoloLens will play a role in surgical education. Furthermore, 80% of participants agreed or strongly agreed that the HoloLens is feasible to be introduced in clinical routine and will play a role within surgery in the future.

Conclusion: The use of fiducial markers and repeated training reduces the positional error between the hologram and the real structures. The developed measuring device under the use of the Fiagon navigation system is suitable to measure accuracies of holographic augmented reality images of the HoloLens.

Keywords: Accuracy augmented reality; Augmented reality; HoloLens; Parotid surgery; Salivary glands.

MeSH terms

  • Augmented Reality*
  • Germany
  • Humans
  • Surgery, Computer-Assisted*