Improving the image quality of DWI in breast cancer: comparison of multi-shot DWI using multiplexed sensitivity encoding to conventional single-shot echo-planar imaging DWI

Br J Radiol. 2021 Mar 1;94(1119):20200427. doi: 10.1259/bjr.20200427. Epub 2020 Sep 9.

Abstract

Objective: To compare diffusion-weighted images (DWI) acquired using single-shot echo-planar imaging (ss-EPI) and multiplexed sensitivity encoding (MUSE) in breast cancer.

Methods: 20 females with pathologically confirmed breast cancer (age 51 ± 12 years) were imaged with ss-EPI-DWI and MUSE-DWI. ADC, normalised ADC (nADC), blur and distortion metrics and qualitative image quality scores were compared. The Crété-Roffet and Mattes mutual information metrics were used to evaluate blurring and distortion, respectively. In a breast phantom, six permutations of MUSE-DWI with varying parallel acceleration factor and number of shots were compared. Differences in ADC and nADC were compared using the coefficient of variation in the phantom and a paired t-test in patients. Differences in blur, distortion and qualitative metrics were analysed using a Wilcoxon signed-rank test.

Results: There was a low coefficient of variation (<2%) in ADC between ss-EPI-DWI and all MUSE-DWI permutations acquired using the phantom. 22 malignant and three benign lesions were identified in 20 patients. ADC values measured using MUSE were significantly lower compared to ss-EPI for malignant but not benign lesions (p < 0.001, p = 0.21). nADC values were not significantly different (p = 0.62, p = 0.28). Blurring and distortion improved with number of shots and acceleration factor, and significantly improved with MUSE in patients (p < 0.001, p = 0.002). Qualitatively, image quality improved using MUSE.

Conclusion: MUSE improves the image quality of breast DWI compared to ss-EPI.

Advances in knowledge: MUSE-DWI has superior image quality and reduced blurring and distortion compared to ss-EPI-DWI in breast cancer.

Publication types

  • Comparative Study

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Breast / diagnostic imaging
  • Breast Neoplasms / diagnostic imaging*
  • Diffusion Magnetic Resonance Imaging / methods*
  • Echo-Planar Imaging / methods*
  • Female
  • Humans
  • Middle Aged
  • Reproducibility of Results
  • Sensitivity and Specificity