Combining Laue diffraction with Bragg coherent diffraction imaging at 34-ID-C

J Synchrotron Radiat. 2020 Sep 1;27(Pt 5):1430-1437. doi: 10.1107/S1600577520009844. Epub 2020 Aug 11.

Abstract

Measurement modalities in Bragg coherent diffraction imaging (BCDI) rely on finding a signal from a single nanoscale crystal object which satisfies the Bragg condition among a large number of arbitrarily oriented nanocrystals. However, even when the signal from a single Bragg reflection with (hkl) Miller indices is found, the crystallographic axes on the retrieved three-dimensional (3D) image of the crystal remain unknown, and thus localizing in reciprocal space other Bragg reflections becomes time-consuming or requires good knowledge of the orientation of the crystal. Here, the commissioning of a movable double-bounce Si (111) monochromator at the 34-ID-C endstation of the Advanced Photon Source is reported, which aims at delivering multi-reflection BCDI as a standard tool in a single beamline instrument. The new instrument enables, through rapid switching from monochromatic to broadband (pink) beam, the use of Laue diffraction to determine crystal orientation. With a proper orientation matrix determined for the lattice, one can measure coherent diffraction patterns near multiple Bragg peaks, thus providing sufficient information to image the full strain tensor in 3D. The design, concept of operation, the developed procedures for indexing Laue patterns, and automated measuring of Bragg coherent diffraction data from multiple reflections of the same nanocrystal are discussed.

Keywords: BCDI; Bragg diffraction; Laue diffraction; strain imaging.