A Preliminary Report of Gonadal-Sparing TBI Using a VMAT Technique

Pract Radiat Oncol. 2021 Mar-Apr;11(2):e134-e138. doi: 10.1016/j.prro.2020.07.006. Epub 2020 Aug 12.

Abstract

Reproductive toxicity is common after total body irradiation (TBI) and has major quality of life implications for patients. In that context, this is the first report of gonadal-sparing volumetric-modulated arc therapy (VMAT) TBI, successfully delivered in a boy and a girl with aplastic anemia. Both patients' VMAT TBI plans demonstrated improved gonadal sparing versus simulated conventional 2-dimensional (2D) approach (mean testes dose, 0.45 Gy VMAT vs 0.72 Gy 2D; mean ovary dose, 0.64 Gy VMAT vs 1.47 Gy 2D). Planning target volume coverage was also improved for both cases with the VMAT plan versus conventional 2D plan (2 Gy D90% vs 1.9 Gy D90%, respectively). Given these dosimetric advantages, the present study can serve as a proof-of-concept for further prospective studies evaluating this technique for wider applications in populations receiving TBI.

MeSH terms

  • Female
  • Humans
  • Male
  • Organs at Risk
  • Prospective Studies
  • Quality of Life
  • Radiotherapy Dosage
  • Radiotherapy Planning, Computer-Assisted
  • Radiotherapy, Intensity-Modulated* / adverse effects
  • Whole-Body Irradiation*