Evaluation of magnetic resonance imaging artifacts caused by fixed orthodontic CAD/CAM retainers-an in vitro study

Clin Oral Investig. 2021 Mar;25(3):1423-1431. doi: 10.1007/s00784-020-03450-x. Epub 2020 Aug 12.

Abstract

Objectives: Magnetic resonance imaging (MRI) image quality can be severely impaired by artifacts caused by fixed orthodontic retainers. In clinical practice, there is a trend towards using computer-aided design/computer-aided manufacturing (CAD/CAM) retainers. This study aimed to quantify MRI artifacts produced by these novel CAD/CAM retainers.

Material and methods: Three CAD/CAM retainers and a stainless-steel retainer ("Twistflex"; clinical reference standard) were scanned in vitro at 3-T MRI using a high-resolution 3D sequence. The artifact diameters and three-dimensional artifact volumes (AV) were determined for all mandibular (AVmand) and maxillary (AVmax) retainers. Moreover, the corresponding ratio of artifact volume to retainer volume (AV/RVmand, AV/RVmax) was calculated.

Results: Twistflex caused large artifact volumes (AVmand: 13530 mm3; AVmax: 15642 mm3; AV/RVmand: 2602; AV/RVmax: 2235). By contrast, artifact volumes for CAD/CAM retainers were substantially smaller: whereas artifact volumes for cobalt-chromium retainers were moderate (381 mm3; 394 mm3; 39; 31), grade-5 titanium (110 mm3; 126 mm3; 12; 12) and nickel-titanium (54 mm3; 78 mm3; 12; 14) both produced very small artifact volumes.

Conclusion: All CAD/CAM retainers caused substantially smaller volumes of MRI artifacts compared to Twistflex. Grade-5 titanium and nickel-titanium CAD/CAM retainers showed the smallest artifact volumes.

Clinical relevance: CAD/CAM retainers made from titanium or nickel-titanium may not relevantly impair image quality in head/neck and dental MRI. Artifacts caused by cobalt-chromium CAD/CAM retainers may mask nearby dental/periodontal structures. In contrast, the large artifacts caused by Twistflex are likely to severely impair diagnosis of oral and adjacent pathologies.

Keywords: Artifacts; CAD/CAM; MRI; Orthodontics; Retainer.