4D in vivo dose verification for real-time tumor tracking treatments using EPID dosimetry

Med Dosim. 2021;46(1):29-38. doi: 10.1016/j.meddos.2020.07.003. Epub 2020 Aug 7.

Abstract

The use of sophisticated techniques such as gating and tracking treatments requires additional quality assurance to mitigate increased patient risks. To address this need, we have developed and validated an in vivo method of dose delivery verification for real-time aperture tracking techniques, using an electronic portal imaging device (EPID)-based, on-treatment patient dose reconstruction and a dynamic anthropomorphic phantom. Using 4DCT scan of the phantom, ten individual treatment plans were created, 1 for each of the 10 separate phases of the respiratory cycle. The 10 MLC apertures were combined into a single dynamic intensity-modulated radiation therapy (IMRT) plan that tracked the tumor motion. The tumor motion and linac delivery were synchronized using an RPM system (Varian Medical Systems) in gating mode with a custom breathing trace. On-treatment EPID frames were captured using a data-acquisition computer with a dedicated frame-grabber. Our in-house EPID-based in vivo dose reconstruction model was modified to reconstruct the 4D accumulated dose distribution for a dynamic MLC (DMLC) tracking plan using the 10-phase 4DCT dataset. Dose estimation accuracy was assessed for the DMLC tracking plan and a single-phase (50% phase) static tumor plan, represented a static field test to verify baseline accuracy. The 3%/3 mm chi-comparison between the EPID-based dose reconstruction for the static tumor delivery and the TPS dose calculation for the static plan resulted in 100% pass rate for planning target volume (PTV) voxels while the mean percentage dose difference was 0.6%. Comparing the EPID-based dose reconstruction for the DMLC tracking to the TPS calculation for the static plan gave a 3%/3 mm chi pass rate of 99.3% for PTV voxels and a mean percentage dose difference of 1.1%. While further work is required to assess the accuracy of this approach in more clinically relevant situations, we have established clinical feasibility and baseline accuracy of using the transmission EPID-based, in vivo patient dose verification for MLC-tracking treatments.

Keywords: EPID dosimetry; Patient dose reconstruction algorithms; Real-time tumor tracking; Treatment verification.

MeSH terms

  • Algorithms
  • Humans
  • Neoplasms* / radiotherapy
  • Phantoms, Imaging
  • Radiometry
  • Radiotherapy Dosage
  • Radiotherapy Planning, Computer-Assisted
  • Radiotherapy, Intensity-Modulated*