Development and Validation of the Elecsys Anti-SARS-CoV-2 Immunoassay as a Highly Specific Tool for Determining Past Exposure to SARS-CoV-2

J Clin Microbiol. 2020 Sep 22;58(10):e01694-20. doi: 10.1128/JCM.01694-20. Print 2020 Sep 22.

Abstract

The Elecsys Anti-SARS-CoV-2 immunoassay (Roche Diagnostics) was developed to provide accurate, reliable detection of antibodies to severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2). We evaluated sensitivity, specificity, cross-reactivity, and agreement with a vesicular stomatitis virus-based pseudoneutralization assay for the Elecsys Anti-SARS-CoV-2 immunoassay. Sensitivity and agreement between Elecsys Anti-SARS-CoV-2 immunoassay and pseudoneutralization assay measurements were evaluated using samples from patients with PCR-confirmed SARS-CoV-2 infection, a majority of whom were hospitalized. Specificity was evaluated using samples from routine diagnostic testing/blood donors collected before December 2019 and thus deemed negative for SARS-CoV-2-specific antibodies. Cross-reactivity was evaluated using samples containing a wide range of potentially cross-reacting analytes, purchased from commercial vendors. For sensitivity and specificity, point estimates and 95% confidence intervals (CIs) were calculated. Agreement between the Elecsys Anti-SARS-CoV-2 immunoassay and the pseudoneutralization assay was calculated. The sensitivity of the Elecsys Anti-SARS-CoV-2 immunoassay in patients with prior PCR-confirmed SARS-CoV-2 infection was 99.5% (95% CI, 97.0 to 100.0%) at ≥14 days post-PCR confirmation. Overall specificity (n = 10,453) was 99.80% (95% CI, 99.69 to 99.88%). Only 4/792 samples containing potential cross-reacting analytes were reactive with the Elecsys Anti-SARS-CoV-2 immunoassay, resulting in an overall specificity in this cohort of 99.5% (95% CI, 98.6 to 99.9%). Positive, negative, and overall agreement (n = 46) between the Elecsys Anti-SARS-CoV-2 immunoassay and the pseudoneutralization assay were 86.4% (95% CI, 73.3 to 93.6%), 100% (95% CI, 34.2 to 100%), and 87.0% (95% CI, 74.3 to 93.9%), respectively. The Elecsys Anti-SARS-CoV-2 immunoassay demonstrated high sensitivity (99.5% at ≥14 days post-PCR confirmation) and specificity (99.80%), supporting its use as a tool for identification of past SARS-CoV-2 infection, including use in populations with low disease prevalence.

Keywords: SARS-CoV-2; cross-reactivity; diagnosis; immunoassay; neutralization testing; sensitivity; specificity.

Publication types

  • Research Support, Non-U.S. Gov't
  • Validation Study

MeSH terms

  • Antibodies, Viral / blood
  • Betacoronavirus / immunology
  • Betacoronavirus / isolation & purification*
  • COVID-19 Testing
  • Clinical Laboratory Techniques
  • Coronavirus Infections / blood
  • Coronavirus Infections / diagnosis
  • Coronavirus Infections / epidemiology
  • Cross Reactions
  • Humans
  • Immunoassay*
  • Reproducibility of Results
  • SARS-CoV-2
  • Sensitivity and Specificity

Substances

  • Antibodies, Viral