The Italian Multicentre Registry of Fenestrated Anaconda™ Endografts for Complex Abdominal Aortic Aneurysms Repair

Eur J Vasc Endovasc Surg. 2020 Aug;60(2):181-191. doi: 10.1016/j.ejvs.2020.04.032. Epub 2020 Jul 21.

Abstract

Objective: The aim was to describe the outcomes of the Anaconda™ Fenestrated endograft Italian Registry for complex aortic aneurysms (AAAs), unsuitable for standard endovascular aneurysm repair (EVAR).

Methods: Between 2012 and 2018 patients with a proximal neck unsuitable for standard EVAR, treated with the fenestrated Anaconda™ endograft, were prospectively enrolled in a dedicated database. Endpoints were peri-operative technical success (TS) and evaluation of type Ia/b or 3 endoleaks (T1/3 EL), target visceral vessel (TVV) occlusion, re-interventions, and AAA related mortality at 30 days, six months, and later follow up.

Results: One hundred twenty seven patients (74 ± 7 years, American Society Anesthesiology (ASA) II/III/IV: 12/85/30) were included in the study in 49 Italian Vascular Surgery Units (83 juxta/para-renal AAA, 13 type IV thoraco-abdominal AAA, 16 T1aEL post EVAR, and 15 short neck AAA). Configurations with one, two, three, and four fenestrations were used in 5, 56, 39, and 27 cases, respectively, for a total of 342 visceral vessels. One hundred and eight (85%) bifurcated and 19 (15%) tube endografts were implanted. In 35% (44/127) of cases the endograft was repositioned during the procedure, and 37% (128/342) of TVV were cannulated from brachial access. TS was 87% (111/127): five T1EL, six T3EL (between fenestration and vessel stent), and six loss of visceral vessels (one patient with a Type Ia EL had also a TVV loss) occurred. Thirty day mortality was 4% (5/127). Two of the five T1EL resolved spontaneously at 30 days. The overall median follow up was 21 ± 16 months; one T1EL (5%) occurred at six months and one T3EL (4%) at the three year follow up. Another two (3%) TVV occlusions occurred at six months and five (3%) at three years. The re-intervention rate at the 30 days, six months, and three year follow up was 5%, 7%, and 18 ± 5%, respectively.

Conclusion: The fenestrated Anaconda™ endograft is effective in the treatment of complex AAA. Some structure properties, such as the re-positionability and the possibility of cannulation from above, are specific characteristics helpful for the treatment of some complex anatomies.

Keywords: Abdominal aortic aneurysm; Anaconda; FEVAR.

Publication types

  • Multicenter Study

MeSH terms

  • Aged
  • Aged, 80 and over
  • Aortic Aneurysm, Abdominal / diagnostic imaging
  • Aortic Aneurysm, Abdominal / mortality
  • Aortic Aneurysm, Abdominal / surgery*
  • Aortic Aneurysm, Thoracic / diagnostic imaging
  • Aortic Aneurysm, Thoracic / mortality
  • Aortic Aneurysm, Thoracic / surgery*
  • Blood Vessel Prosthesis Implantation / adverse effects
  • Blood Vessel Prosthesis Implantation / instrumentation*
  • Blood Vessel Prosthesis Implantation / mortality
  • Blood Vessel Prosthesis*
  • Endoleak / etiology
  • Endoleak / therapy
  • Endovascular Procedures / adverse effects
  • Endovascular Procedures / instrumentation*
  • Endovascular Procedures / mortality
  • Female
  • Humans
  • Italy
  • Male
  • Prospective Studies
  • Prosthesis Design
  • Registries
  • Risk Factors
  • Time Factors
  • Treatment Outcome