A Swine Model of Filter-Assisted Caval Thrombosis for Endovascular Thrombectomy Using AngioJet

J Cardiovasc Transl Res. 2021 Apr;14(2):365-370. doi: 10.1007/s12265-020-10059-8. Epub 2020 Jul 13.

Abstract

To assess the feasibility of creating swine model of filter-assisted caval thrombosis and to evaluate the efficacy in removing clot in this model using rheolytic thrombectomy. The model was created by implanting a filter into the inferior vena cava followed by injection of autologous thrombus. Rheolytic thrombectomy was performed for all models to remove the clot. The success rate of model creation and the efficacy of clot removal were analyzed. The success rate of model creation was 100% (15/15). Following rheolytic thrombectomy, 3 of 5 pigs attained complete clot removal in a 7-day-old model, while no pigs attained complete clot removal in 14- and 21-day-old models. Creating a filter-related caval thrombosis model in swine is technically feasible and can be used to mimic a clinical episode of caval thrombosis from acute phase to chronic occlusion. Rheolytic thrombectomy can be used to remove filter-related thrombosis that aged less than 14 days. Graphical Abstract Swine Model of Filter-Assisted Caval Thrombotic Occlusion.

Keywords: Animal model; Filter-related thrombosis; Inferior vena cava; Rheolytic thrombectomy.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Disease Models, Animal
  • Endovascular Procedures / instrumentation*
  • Feasibility Studies
  • Phlebography
  • Sus scrofa
  • Thrombectomy / instrumentation*
  • Time Factors
  • Vena Cava Filters*
  • Vena Cava, Inferior* / diagnostic imaging
  • Vena Cava, Inferior* / pathology
  • Venous Thrombosis / diagnostic imaging
  • Venous Thrombosis / etiology
  • Venous Thrombosis / pathology
  • Venous Thrombosis / therapy*