Comparison of Novel Immunoassay With Liquid Chromatography/Tandem Mass Spectrometry (LC-MS/MS) for Therapeutic Drug Monitoring of Clozapine

Ther Drug Monit. 2020 Oct;42(5):771-777. doi: 10.1097/FTD.0000000000000777.

Abstract

Background: Clozapine is the most effective antipsychotic for treatment-resistant schizophrenia. Although serum clozapine levels can help guide treatment, they are underutilized owing to requirements for frequent venous blood draws and lack of immediate results.

Methods: Clozapine levels measured with a novel immunoassay technology (which enables point-of-care development) were compared with those measured by standard liquid chromatography/tandem mass spectrometry (LC-MS/MS). Frozen serum aliquots of 117 samples (N = 48 patients with schizophrenia on clozapine; N = 24 patients with schizophrenia not on clozapine; N = 45 healthy controls) were sent to a national reference laboratory (NRL) for clozapine level determination by LC-MS/MS, and matching samples were subjected to novel immunoassay (3 runs). At a later date, another frozen aliquot from the same date was sent to the NRL for repeat testing.

Results: The NRL obtained 18 false-positive clozapine results (mean 42.39 ± 32.06, range 21-159 ng/mL) in participants not on clozapine (N = 3) and healthy controls (N = 15). The immunoassay showed no false-positive clozapine results. The clozapine levels were correlated between both assays (r = 0.84, P < 0.0001), despite 16% higher clozapine levels with immunoassay (482.08 ± 270.88 ng/mL immunoassay, 414.98 ± 186.29 ng/mL LC-MS/MS [P = 0.03]). Agreement analysis using concordance correlation coefficient (CCC) for LC-MS/MS of the 2 aliquots yielded CCC = 0.869; 95% confidence interval = 0.690-0.970, whereas higher agreement results were observed for the 3 runs of immunoassay (CCC = 0.99; 95% confidence interval = 0.979-0.997).

Conclusions: The lack of false positives observed with immunoassay, higher repeat performance agreement, and good correlation with LC-MS/MS may indicate the more robust performance of immunoassay than that of LC-MS/MS clozapine-level determination.

Publication types

  • Comparative Study

MeSH terms

  • Adult
  • Antipsychotic Agents / blood
  • Antipsychotic Agents / therapeutic use
  • Chromatography, Liquid / methods*
  • Clozapine / blood*
  • Clozapine / therapeutic use
  • Drug Monitoring / methods*
  • Female
  • Humans
  • Immunoassay / methods*
  • Male
  • Middle Aged
  • Point-of-Care Systems
  • Schizophrenia / blood
  • Schizophrenia / drug therapy
  • Tandem Mass Spectrometry / methods*
  • Young Adult