Videofluoroscopic Evaluation of the Influence of a Gradually Reducing Femoral Radius on Joint Kinematics During Daily Activities in Total Knee Arthroplasty

J Arthroplasty. 2020 Oct;35(10):3010-3030. doi: 10.1016/j.arth.2020.05.039. Epub 2020 May 27.

Abstract

Background: Paradoxical anterior translation in midflexion is reduced in total knee arthroplasties (TKAs) with a gradually reducing femoral radius, when compared to a 2-radii design. This reduction has been shown in finite element model simulations, in vitro tests, intraoperatively, and recently also in vivo during a lunge and unloaded flexion-extension. However, TKA kinematics are task dependent and this reduction has not been tested for gait activities.

Methods: Thirty good outcome subjects (≥1 year postoperatively) with a unilateral cruciate-retaining TKA with a gradually reducing (n = 15) or dual (n = 15) femoral radius design were assessed during 5 complete cycles of level walking, stair descent (0.18-m steps), deep knee bend, and sitting down onto and standing up from a chair, using a moving fluoroscope (25 Hz, 1 ms shutter time). Kinematic data were extracted by 2D/3D image registration.

Results: Tibiofemoral ranges of motion for flexion-extension, abduction-adduction, internal-external rotation, and anteroposterior (AP) translation were similar for both groups, whereas the pattern of AP translation-flexion-coupling differed. The subjects with the dual-radii design showed a sudden change in direction of AP translation around 30° of flexion, which was not present in the subjects with the gradually reducing femoral radius design.

Conclusion: Through the unique ability of moving fluoroscopy, the present study confirmed that the gradually reducing femoral radii eliminated the paradoxical sudden anterior translation at 30° present in the dual-radii design in vivo during daily activities, including gait and stair descent.

Keywords: Attune; Sigma; condylar geometry; gait analysis; in vivo kinematics; moving fluoroscopy.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Arthroplasty, Replacement, Knee*
  • Biomechanical Phenomena
  • Humans
  • Knee Joint / diagnostic imaging
  • Knee Joint / surgery
  • Knee Prosthesis*
  • Radius
  • Range of Motion, Articular