Intraoperative Neuromonitoring Using a Single Transcartilage Needle Electrode During Thyroidectomy

Laryngoscope. 2021 Feb;131(2):448-452. doi: 10.1002/lary.28835. Epub 2020 Jun 20.

Abstract

Objectives: Application of transcartilage needle electrode for intraoperative neuromonitoring (IONM) during thyroidectomy has been considered an alternative method of electromyography (EMG) tube recording. However, needle electrodes must be inserted into both sides of the thyroid cartilage with exposure of the cartilage lamina. We sought to evaluate the feasibility of applying a single ipsilateral transcartilage needle electrode for IONM during unilateral hemithyroidectomy.

Methods: Thirty-four patients underwent IONM during unilateral thyroidectomy. A dual disposable needle electrode was inserted obliquely into the inferior lower third of the ipsilateral lamina of the thyroid cartilage. Patients were classified as deep (≥ 5 mm) or superficial (< 5 mm) by the depth of the needle electrode inserted into the thyroid cartilage. Without using an EMG tube, IONM was done according to the standardized procedure using a single needle electrode only.

Results: IONM was successful in all nerves at risk. Amplitude of the EMG signal was stable during the surgery, with no cases presenting loss of signal. Amplitude of the signal from vagal and recurrent laryngeal nerve stimulation was significantly lower when needle insertion was superficial (< 5 mm). None of the patients showed postoperative vocal cord paralysis, and complications related to needle electrode were not identified.

Conclusion: IONM using a single ipsilateral transcartilage needle electrode during unilateral hemithyroidectomy was feasible with no significant complications. It may serve as an alternative method of IONM with the advantage of minimal exposure of the thyroid cartilage lamina.

Level of evidence: 4 Laryngoscope, 131:448-452, 2021.

Keywords: Thyroidectomy, recurrent laryngeal nerve, neuromonitoring, needle electrode.

Publication types

  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Aged
  • Electrodes
  • Equipment Design
  • Feasibility Studies
  • Female
  • Humans
  • Intraoperative Neurophysiological Monitoring / instrumentation*
  • Intraoperative Neurophysiological Monitoring / methods
  • Male
  • Middle Aged
  • Needles
  • Postoperative Complications / etiology
  • Postoperative Complications / prevention & control*
  • Recurrent Laryngeal Nerve / surgery
  • Recurrent Laryngeal Nerve Injuries / etiology
  • Recurrent Laryngeal Nerve Injuries / prevention & control
  • Thyroid Cartilage / innervation
  • Thyroid Cartilage / surgery*
  • Thyroidectomy / adverse effects
  • Thyroidectomy / methods*
  • Treatment Outcome
  • Vagus Nerve / surgery
  • Vocal Cord Paralysis / etiology
  • Vocal Cord Paralysis / prevention & control*
  • Young Adult