Implant Diameter, Length, and the Insertion Torque/Depth Integral: A Study Using Polyurethane Foam Blocks

Dent J (Basel). 2020 Jun 4;8(2):56. doi: 10.3390/dj8020056.

Abstract

The amount of energy necessary to place an implant in its seat, described as the integral of the torque-depth curve at insertion (I), has been validated as a reliable measure of primary stability. This study aimed to investigate whether (I) may detect the variations in primary stability caused by changes in the implant length or diameter better than the insertion torque (IT). Cylindric implants featuring a double-etched, sandblasted surface with different diameters or lengths were placed into monolithic polyurethane foam blocks with different densities that mimicked human bone. (I)-, (I)*-, IT-, IT*-diameter and -length plots ((I)* and IT* were the derived values corrected for undersizing) were drawn and the relation between (I), (I)*, IT, and IT* and the fixture diameter or length was investigated with correlation analysis. (I)* and IT* correlated better than (I) and IT with the fixture diameter; (I), (I)*, IT, and IT* correlated equally well with the fixture length. In all cases, the slopes of the lines best fitting the experimental data were greater for (I) or (I)* than IT or IT*, respectively. (I) or (I)* were better detectors than IT or IT* of the changes in primary stability that can be achieved by increasing the fixture diameter or length.

Keywords: fixture diameter; fixture length; insertion torque; primary stability; torque-depth curve integral; undersizing.