Comparison and Combination of Single-Lead ECG and Photoplethysmography Algorithms for Wearable-Based Atrial Fibrillation Screening

Telemed J E Health. 2021 Mar;27(3):296-302. doi: 10.1089/tmj.2020.0036. Epub 2020 May 18.

Abstract

Background:Atrial fibrillation (AF), the most common cardiac arrhythmia, can be detected by smartphones and smartwatches.Introduction:Single-lead ECGs (iECGs) and photoplethysmography (PPG) sensors provide the opportunity for a broad, simple, and easily repeatable cardiac rhythm analysis. To reduce unnecessary medical follow-up testing due to false positive results, our aim was to find a screening approach applicable on smart devices with a focus on high specificity.Methods:We used PPG measurements from smartphones and smartwatches and iECG data from two previous validation trials. Two AF detection algorithms (A and B) were applied on the iECG dataset and compared directly. Further, we used 1-min PPG measurements as a first-pass filter for arrhythmia detection and simulated a sequential testing: Once an arrhythmia was detected in the PPG, the iECG counterpart of the patient was analyzed by algorithm A, B, or A + B combined although algorithm B was primarily designed for PPG analysis.Results:The iECGs from 1,288 participants were analyzed. Algorithm A did not show a diagnosis in 16.1%. In the remaining, sensitivity and specificity were 99.6%, and 97.4% respectively. Accuracy was 98.5%, and correct classification rate (CCR) was 82.7%. Algorithm B always differentiated between normal and arrhythmic and reached an overall sensitivity of 95.4%, a specificity of 91.6%, and an accuracy and CCR of 93.3%. Sequential testing by combining both algorithms into a three-phase test (Test positive PPG, then iECG analysis by A and B combined) resulted in a 100% specificity.Conclusion:Algorithm B performed strongly in PPG analysis as well as iECG analysis. PPG signals and consecutive iECG combined when an arrhythmia was detected by PPG resulted in a specificity that was higher than 99%.Discussion:The analysis allows a direct comparison of iECG algorithms without possible dilution by different measurement procedures or recording-devices. We improved specificity in AF-screening approaches with wearables by simulating a novel approach. Results rely on signal quality.

Keywords: atrial fibrillation; eHealth; iECG; photoplethysmography; smartphone; wearables.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Algorithms
  • Atrial Fibrillation* / diagnosis
  • Electrocardiography
  • Heart Rate
  • Humans
  • Photoplethysmography
  • Prospective Studies
  • Wearable Electronic Devices*