Comparison of objective and subjective visual quality early after implantable collamer lens V4c (ICL V4c) and small incision lenticule extraction (SMILE) for high myopia correction

Acta Ophthalmol. 2020 Dec;98(8):e943-e950. doi: 10.1111/aos.14459. Epub 2020 May 18.

Abstract

Purpose: To investigate differences in the objective and subjective visual quality of patients with high myopia following implantable collamer lens V4c (ICL V4c) versus small incision lenticule extraction (SMILE) early postoperatively.

Methods: This prospective analysis included 94 eyes of 57 patients (8 males) in the ICL V4c group (mean spherical equivalent (SE), -8.07 ± 1.03 dioptres (D)) and 103 eyes of 57 patients (13 males) in the SMILE group (mean SE, -7.85 ± 0.78 D). Refractive parameters and wavefront aberrations were recorded preoperatively and 6 months postoperatively. Subjective visual quality was scored with the Quality of Vision questionnaire, and operation satisfaction was evaluated 6 months postoperatively.

Results: At 6 months postoperatively, the efficacy index was 1.17 ± 0.16 in the ICL and 1.09 ± 0.20 in the SMILE group; the safety index was 1.21 ± 0.18 in the ICL and 1.12 ± 0.16 in the SMILE group. A significant increase in total higher-order aberrations (HOAs) in both the ICL and SMILE groups was observed, but to a lesser degree in ICL (p < 0.05). In the ICL group, trefoil significantly increased (p < 0.05), whereas in the SMILE group, coma, trefoil and spherical aberration significantly increased (p < 0.05). Glare, haloes and blurred vision had the highest incidences after both types of surgery, and postoperative haloes were significantly more common following ICL; 96.1% of ICL-treated and 94.9% of SMILE-treated patients were satisfied with their visual outcomes.

Conclusion: ICL V4c caused lower HOAs induction than SMILE. Despite the postoperative visual disturbances, both ICL and SMILE provided good efficacy, safety, predictability and high satisfaction in correcting high myopia.

Keywords: high myopia; implantable collamer lens V4c (ICL V4c); satisfaction; small incision lenticule extraction (SMILE); visual quality; wavefront aberration.

MeSH terms

  • Adult
  • Corneal Surgery, Laser / methods*
  • Corneal Topography
  • Female
  • Follow-Up Studies
  • Humans
  • Lasers, Excimer / therapeutic use*
  • Lenses, Intraocular*
  • Male
  • Middle Aged
  • Myopia / physiopathology
  • Myopia / surgery*
  • Postoperative Period
  • Prospective Studies
  • Prosthesis Design
  • Refraction, Ocular / physiology*
  • Visual Acuity*
  • Young Adult