Posterior fossa ICP monitoring: a tale of two compartments

Br J Neurosurg. 2021 Apr;35(2):129-132. doi: 10.1080/02688697.2020.1765974. Epub 2020 May 15.

Abstract

Aim: Measuring the intracranial pressure (ICP) of the infra-tentorial, posterior fossa compartment has long been avoided due to a lack of precedent and interpretability, as well as concern of damage to the underlying vital structures. In cases of posterior fossa insults however, the supra-tentorial compartment ICPs can be falsely reassuring. We aimed to measure the posterior fossa ICP in such a case and analyse the resulting data.

Methods: We present a case of posterior fossa ICP monitoring and discuss its safety profile, rationale and possible indications.

Results: Our comparison of the supra and infra-tentorial ICPs showed that there was a statistically significant difference in the two compartments. The infra-tentorial compartment had ICPs averaging 11.02 ± 2.24 mmHg whilst the supra-tentorial compartment averaged 4.94 ± 1.80 mmHg in the first 72 hours post-op (p < .01 on paired t-testing). After 72 hours, the pressures seemed to equilibrate and were 4.71 ± 2.6 and 3.88 ± 2.89 for the infra and supra-tentorial compartments respectively.

Conclusion: We propose that where a patient with a posterior fossa insult exhibits signs and symptoms consistent with raised ICP but the supra-tentorial readings are normal, posterior fossa ICP monitoring can be considered.

Keywords: ICP monitoring; Posterior fossa; neuro-critical care.

Publication types

  • Case Reports

MeSH terms

  • Dura Mater
  • Humans
  • Intracranial Pressure*
  • Monitoring, Physiologic
  • Skull*