Traceable point-like 68Ge/68Ga source with a spherically symmetric positron absorber for PET scanners

Radiol Phys Technol. 2020 Jun;13(2):170-176. doi: 10.1007/s12194-020-00565-4. Epub 2020 May 2.

Abstract

A novel traceable point-like 68Ge/68Ga source was developed for calibration of positron emission tomography (PET) scanners. The source was equipped with a spherically symmetric acrylic positron absorber. The physical characteristics of photons emitted from the point-like source were evaluated by Monte Carlo simulation, considering possible geometrical uncertainties. The calibration factor of a clinical PET/CT scanner was evaluated using four manufactured point-like sources as a practical application of the point-like source. The results were consistent with that determined by the conventional cross-calibration method. The traceable point-like 68Ge/68Ga source is expected to be a simple and practical tool for determining the calibration factor and evaluating the physical characteristics of PET scanners.

Keywords: Calibration; PET; Point-like source; Quality control.