The effect of stereotactic body radiotherapy (SBRT) using flattening filter-free beams on cardiac implantable electronic devices (CIEDs) in clinical situations

J Appl Clin Med Phys. 2020 Jun;21(6):121-131. doi: 10.1002/acm2.12873. Epub 2020 Apr 11.

Abstract

Purpose: This study focused on determining risks from stereotactic radiotherapy using flattening filter-free (FFF) beams for patients with cardiac implantable electronic device (CIEDs). Two strategies were employed: a) a retrospective analysis of patients with CIEDs who underwent stereotactic radiosurgery (SRS)/SBRT at the Peter MacCallum Cancer Centre between 2014 and 2018 and b) an experimental study on the impact of FFF beams on CIEDs.

Methods: A retrospective review was performed. Subsequently, a phantom study was performed using 30 fully functional explanted CIEDs from two different manufacturers. Irradiation was carried out in a slab phantom with 6-MV and 10-MV FFF beams. First, a repetition-rate test (RRT) with a range of beam pulse frequencies was conducted. Then, multifraction SBRT (48 Gy/4 Fx) and single-fraction SBRT (28 Gy/1 Fx) treatment plans were used for lung tumors delivered to the phantom.

Results: Between 2014 and 2018, 13 cases were treated with an FFF beam (6 MV, 1400 MU/min or 10 MV, 2400 MU/min), and 15 cases were treated with a flattening filter (FF) beam (6 MV, 600 MU/min). All the devices were positioned outside the treatment field at a distance of more than 5 cm, except for one case, and no failures were reported due to SBRT/SRS. In the phantom rep-rate tests, inappropriate sensing occurred, starting at a rep-rate of 1200 MU/min. Cardiac implantable electronic device anomalies during and after delivering VMAT-SBRT with a 10-MV FFF beam were observed.

Conclusions: The study showed that caution should be paid to managing CIED patients when they undergo SBRT using FFF beams, as it is recommended by AAPM TG-203. Correspondingly, it was found that for FFF beams although there is small risk from dose-rate effects, delivering high dose of radiation with beam energy greater than 6 MV and high-dose rate to CIEDs positioned in close vicinity of the PTV may present issues.

Keywords: cardiac implantable electronic device (CIED); flattening filter-free (FFF) beams; stereotactic body radiotherapy (SBRT); volumetric-modulated arc therapy (VMAT).

MeSH terms

  • Electronics
  • Humans
  • Radiosurgery*
  • Radiotherapy Dosage
  • Radiotherapy Planning, Computer-Assisted
  • Radiotherapy, Intensity-Modulated*
  • Retrospective Studies