Accurate integral counting using multi-channel analyzers

Appl Radiat Isot. 2020 May:159:109101. doi: 10.1016/j.apradiso.2020.109101. Epub 2020 Feb 27.

Abstract

Many techniques in radionuclide metrology rely on accurate measurement of the total count rate, that is integral counting, from a detector. We present the design and results from integral counting using a live-timed multi-channel analyzer (MCA) based on a list-mode digitizer and a live-timed scalar, both with imposed, extending dead times. These systems were benchmarked against an existing NIST anti-coincidence counting system. Both systems were used to determine correction factors to an MCA that uses Gedcke-Hale dead-time corrections. Tests were performed for a large-area gas-flow proportional counter and a NaI(Tl) detector for count rates up to 80 000 s-1. The results were used to create correction factors suitable for alpha- and beta-particle emission rate calibrations.

Keywords: Dead time; Integral counting; MCA; Multi-channel analyzer; Radionuclide metrology.