Purpose: To assess the correlation between postimplant dosimetric quantifiers and the genitourinary (GU) toxicity of low-dose rate brachytherapy for prostate cancer.
Methods and materials: The minimum urethral dose (UD10, 30, and 90) and the percent volume of the urethra receiving the prescription dose (V100, V150) were calculated from the postimplant dose-volume histograms of 182 patients. We then calculated various urethral biologically equivalent doses (uBEDs) using different values of the α/β ratio and tissue repair half-time (t1/2) and examined the correlations with GU toxicity.
Results: Common dosimetric quantifiers, such as UD90 (brachytherapy) + UD50 (external beam radiotherapy), showed no correlation with Grade ≥ 2 GU toxicity. There was a significant correlation between Grade ≥2 GU toxicity and uBED when the α/β value was 0.5 or 1 Gy and t1/2 was 0.5-2.5 h. An uBED (α/β = 1.0, t1/2 = 0.5) had the largest hazard ratio for GU toxicity, and it was also significantly correlated with Grade ≥ 2 GU toxicity according to multivariate analysis.
Conclusions: We observed a significant correlation of uBED with GU toxicity when α/β was 0.5 or 1.0 Gy and t1/2 was 0.5-2.5 h. As the simple formula we used has not been verified in basic experiments, more data are needed to validate our results.
Keywords: Biologically equivalent dose; Late genitourinary toxicity; Low-dose-rate brachytherapy; Prostate cancer; Tissue repair half-time; α/β ratio.