Non-contact infrared versus axillary and tympanic thermometers in children attending primary care: a mixed-methods study of accuracy and acceptability

Br J Gen Pract. 2020 Mar 26;70(693):e236-e244. doi: 10.3399/bjgp20X708845. Print 2020 Apr.

Abstract

Background: Guidelines recommend measuring temperature in children presenting with fever using electronic axillary or tympanic thermometers. Non-contact thermometry offers advantages, yet has not been tested against recommended methods in primary care.

Aim: To compare two different non-contact infrared thermometers (NCITs) to axillary and tympanic thermometers in children aged ≤5 years visiting their GP with an acute illness.

Design and setting: Method comparison study with nested qualitative component.

Method: Temperature measurements were taken with electronic axillary (Welch Allyn SureTemp®), electronic tympanic (Braun Thermoscan®), NCIT Thermofocus® 0800, and NCIT Firhealth Forehead. Parents rated acceptability and discomfort. Qualitative interviews explored parents' experiences of the thermometers.

Results: In total, 401 children were recruited (median age 1.6 years, 50.62% male). Mean difference between the Thermofocus NCIT and axillary thermometer was -0.14°C (95% confidence interval [CI] = -0.21 to -0.06°C); lower limit of agreement was -1.57°C (95% CI = -1.69 to -1.44°C) and upper limit 1.29°C (95% CI = 1.16 to 1.42°C). A second NCIT (Firhealth) had similar levels of agreement; however, the limits of agreement between tympanic and axillary thermometers were also wide. Parents expressed a preference for the practicality and comfort of NCITs, and were mostly negative about their child's experience of axillary thermometers. But there was willingness to adopt whichever device was medically recommended.

Conclusion: In a primary care paediatric population, temperature measurements with NCITs varied by >1°C compared with axillary and tympanic approaches. But there was also poor agreement between tympanic and axillary thermometers. Since clinical guidelines often rely on specific fever thresholds, clinicians should interpret peripheral thermometer readings with caution and in the context of a holistic assessment of the child.

Keywords: acute disease; child; fever; primary health care; thermometers.

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Axilla
  • Body Temperature
  • Child, Preschool
  • Cross-Sectional Studies
  • Female
  • Fever / diagnosis*
  • Forehead
  • Humans
  • Infant
  • Infrared Rays
  • Male
  • Patient Preference
  • Primary Health Care*
  • Sensitivity and Specificity
  • Thermometers*
  • Tympanic Membrane