Peristaltic Contrast Media Injection Improved Image Quality and Decreased Radiation and Contrast Dose When Compared With Direct Drive Injection During Liver Computed Tomography

J Comput Assist Tomogr. 2020 Mar/Apr;44(2):209-216. doi: 10.1097/RCT.0000000000000994.

Abstract

Purpose: The aim of this study was to compare hepatic vascular and parenchymal image quality between direct and peristaltic contrast injectors during hepatic computed tomography (HCT).

Methods: Patients (n = 171) who underwent enhanced HCT and had both contrast media protocols and injector systems were included; group A: direct-drive injector with fixed 100 mL contrast volume (CV), and group B: peristaltic injector with weight-based CV. Opacification, contrast-to-noise ratio, signal-to-noise ratio, radiation dose, and CV for liver parenchyma and vessels in both groups were compared by paired t test and Pearson correlation. Receiver operating characteristic curve, visual grading characteristics, and Cohen κ were used.

Results: Contrast-to-noise ratio: compared with hepatic vein for functional liver, contrast-to-noise ratio was higher in group B (2.17 ± 0.83) than group A (1.82 ± 0.63); portal vein: higher in group B (2.281 ± 0.96) than group A (2.00 ± 0.66). Signal-to-noise ratio for functional liver was higher in group B (5.79 ± 1.58 Hounsfield units) than group A (4.81 ± 1.53 Hounsfield units). Radiation dose and contrast media were lower in group B (1.98 ± 0.92 mSv) (89.51 ± 15.49 mL) compared with group A (2.77 ± 1.03 mSv) (100 ± 1.00 mL). Receiver operating characteristic curve demonstrated increased reader in group B (95% confidence interval, 0.524-1.0) than group A (95% confidence interval, 0.545-1.0). Group B had increased revenue up to 58% compared with group A.

Conclusions: Image quality improvement is achieved with lower CV and radiation dose when using peristaltic injector with weight-based CV in HCT.

Publication types

  • Comparative Study

MeSH terms

  • Contrast Media / administration & dosage*
  • Female
  • Humans
  • Injections, Intravenous
  • Liver / diagnostic imaging*
  • Male
  • Middle Aged
  • Radiation Dosage*
  • Radiographic Image Enhancement / methods*
  • Retrospective Studies
  • Tomography, X-Ray Computed / methods*

Substances

  • Contrast Media