Optimizing the Calibration Error of Refraction Angles in Ultrasonic Angle Beam Testing

Sensors (Basel). 2020 Mar 5;20(5):1427. doi: 10.3390/s20051427.

Abstract

Ultrasonic testing is a useful approach for quantifying the flaws in mechanical components. The height of the flaws in ultrasonic angle beam testing is closely related to the calibration value of the probe refraction angle. In order to reduce the calibration error, some ignored data during the traditional calibration process are reanalyzed and fused to determine the refraction angle. Both arithmetical measurement fusion method and weighted measurement fusion method are applied and compared. Monte Carlo simulation is used to estimate the probability distribution of the refraction angle and obtain the optimal refraction angle weights. Experiments were carried out to verify the results of Monte Carlo simulation. The applicability of data fusion on refraction angles is investigated. It was found in the study that the data fused with the refraction angle is helpful for measuring the height of flaws.

Keywords: data fusion; instrument calibration; optimal weights; ultrasonic angle beam; ultrasonic testing; weighted measurement.