A Disposable Passive Microfluidic Device for Cell Culturing

Biosensors (Basel). 2020 Feb 29;10(3):18. doi: 10.3390/bios10030018.

Abstract

In this work, a disposable passive microfluidic device for cell culturing that does not require any additional/external pressure sources is introduced. By regulating the height of fluidic columns and the aperture and closure of the source wells, the device can provide different media and/or drug flows, thereby allowing different flow patterns with respect to time. The device is made of two Polymethylmethacrylate (PMMA) layers fabricated by micro-milling and solvent assisted bonding and allows us to ensure a flow rate of 18.6 μl/ℎ - 7%/day, due to a decrease of the fluid height while the liquid is driven from the reservoirs into the channels. Simulations and experiments were conducted to characterize flows and diffusion in the culture chamber. Melanoma tumor cells were used to test the device and carry out cell culturing experiments for 48 hours. Moreover, HeLa, Jurkat, A549 and HEK293T cell lines were cultivated successfully inside the microfluidic device for 72 hours.

Keywords: Cell Culture; Micro-Bioreactor; Passive Microfluidic Device.

MeSH terms

  • Cell Culture Techniques / methods*
  • Humans
  • Microfluidics / methods*