Near-Infrared Tunable Laser Absorption Spectroscopic Acetylene Sensor System Using a Novel Three Mirror-Based, Dense Pattern Gas Cell

Sensors (Basel). 2020 Feb 26;20(5):1266. doi: 10.3390/s20051266.

Abstract

By contrast with the widely reported traditional two mirror-based Herriott cell, a three mirror-based dense pattern gas cell was proposed, of which the modeling and design were proven to be effective through a comparison between the simulated spot pattern and effective path length and those of the experimental results. A mechanical structure was designed to adjust the position/angle of the three mirrors for aligning the optical path. The experimentally measured reflection number was 60, resulting in an optical path length of ~11 m, which agrees well with the theoretical value of 10.95 m. Combined with a near-infrared laser with a center wavenumber located at an acetylene (C2H2) absorption line of 6521.2 cm-1, a C2H2 sensor system was established to verify the feasibility of the three mirror-based gas cell. Assisted by a data acquisition (DAQ) card, a LabVIEW platform was developed to generate the drive signal of the laser and acquire the second harmonic (2f) signal from the output of the detector. Through Allan variance analysis, the limit of detection (LoD) of the sensor system is 4.36 ppm at an average time of 0.5 s; as the average time exceeds 10 s, the LoD is <1 ppm. The proposed model and design of the three mirror-based gas cell can be used to realize similar gas cells with different absorption path lengths for gas detection based on infrared absorption spectroscopy.

Keywords: acetylene detection; laser absorption spectroscopy; multipass gas cell; wavelength modulation spectroscopy.