Demands for mechanical accuracy of medical linear accelerators are increased due to the stereotactic and modulated rotational treatments. Mechanical inaccuracies affect the size and shape of the mechanical and radiation isocenters. In practice, the mechanical isocenter is defined by the intersection of rotational axes. However, there are no simple tools to check the properties of the mechanical isocenter in 3D. We introduce a new photography-based method for quick and sub-millimeter accurate determination of the mechanical isocenter. The method is based on image-processing algorithm and modified front pointer. The results demonstrate the quick measurement and visualization of the mechanical isocenter.
Keywords: Isocenter; Linear accelerator; Mechanical; Quality assurance (QA).