Measuring the dose-width product and proposing the local diagnostic reference level in panoramic dental radiography: a multi-center study from Iran

Oral Radiol. 2021 Jan;37(1):80-85. doi: 10.1007/s11282-020-00428-z. Epub 2020 Feb 17.

Abstract

Objective: Although radiation exposure associated with dental radiography is relatively low, patient exposure must be kept practically low. Therefore, it is necessary for each country to establish its own diagnostic reference levels (DRLs) suitable for its equipment and practice. In the present study, dose-width product (DWP) values for panoramic dental radiography were measured and a local DRL was established.

Methods: Five panoramic devices from five radiology clinics of Kashan, Iran were selected to measure the DWP values of panoramic dental radiography. To investigate the DWP values, the parameters of each patient's exposure (e.g., tube voltage, tube current, and exposure time) at these five radiology clinics were extracted. Then, the dose value received by each patient was measured based on a CT pencil chamber. Finally, the overall median DWP values for the patients with small, medium, and large sizes were obtained, and these values were considered as the local DRLs for panoramic dental radiography.

Results: A total of 99 adult patients were included in the present study. The findings demonstrated that the median and third-quartile DWP values for these five radiology clinics ranged from 42.3 to 94.3 and 49.7 to 142.8 mGy mm, respectively. The local DRL values, which were established as the overall median DWP values, were 43.4, 52.0, and 80.3 mGy mm for the adults with small, medium, and large sizes, respectively.

Conclusion: The local DRL proposed in this study for the adult with standard/medium size was lower than those proposed by other reports and seemed acceptable for panoramic radiography in Kashan, Iran.

Keywords: Diagnostic reference level (DRL); Dose–width product (DWP); Iran; Panoramic radiography.

Publication types

  • Multicenter Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Diagnostic Reference Levels*
  • Humans
  • Iran
  • Radiation Dosage
  • Radiography, Panoramic
  • Radiometry*