Trifecta Versus Perimount Magna Ease Aortic Valve Prostheses

Ann Thorac Surg. 2020 Sep;110(3):879-888. doi: 10.1016/j.athoracsur.2019.12.071. Epub 2020 Feb 12.

Abstract

Background: Recent surgical bovine pericardial prostheses are widely used despite limited data on their long-term durability.

Methods: This is a comparative analysis of the outcome of the Trifecta (Abbott, Abbott Park, IL) and Perimount Magna Ease (Edwards, Lifesciences Corp, Irvine, CA) bioprostheses from the FinnValve registry, a Finnish nationwide database including patients with aortic stenosis who underwent transcatheter or surgical aortic valve replacement with a bioprosthesis between 2008 and 2017.

Results: Overall 2216 patients (mean age, 74.1 ± 6.7 years; age <65 years, 8.9%; mean follow-up, 3.8 ± 2.1 years) received the Trifecta (851 patients) or the Perimount Magna Ease (1365 patients) bioprostheses. The rates of late mortality and prosthetic valve endocarditis were comparable in the study cohorts. At 7 years the Trifecta cohort had a significantly higher risk of repeat aortic valve replacement for structural valve failure (3.3% vs 0%; adjusted subdistribution hazard ratio, 2.224; 95% confidence interval, 1.044-4.737), repeat aortic valve replacement for any cause (3.6% vs 0.4%; adjusted subdistribution hazard ratio, 3.210; 95% confidence interval, 1.286-8.013), and repeat aortic valve replacement and/or prosthetic valve endocarditis (4.1% vs 0.9%; adjusted subdistribution hazard ratio, 3.210; 95% confidence interval, 1.286-8.013) compared with the Perimount Magna Ease cohort. Among 772 propensity score-matched pairs, at 7 years the Trifecta cohort had a higher risk of repeat aortic valve replacement for structural valve failure (5.7% vs 0%, P = .009).

Conclusions: The Trifecta aortic bioprosthesis is associated with a higher occurrence of repeat aortic valve replacement for structural valve failure compared with the Perimount Magna Ease bioprosthesis. Further comparative studies with echocardiographic data on structural valve deterioration and longer follow-up are needed to confirm these findings.

Publication types

  • Comparative Study
  • Multicenter Study

MeSH terms

  • Aged
  • Aortic Valve / diagnostic imaging
  • Aortic Valve / surgery*
  • Aortic Valve Stenosis / diagnosis
  • Aortic Valve Stenosis / surgery*
  • Bioprosthesis*
  • Echocardiography
  • Female
  • Follow-Up Studies
  • Heart Valve Prosthesis
  • Heart Valve Prosthesis Implantation
  • Humans
  • Male
  • Prosthesis Design
  • Retrospective Studies
  • Time Factors