Accelerating availability of clinically-relevant parameter estimates from thromboelastogram point-of-care device

J Trauma Acute Care Surg. 2020 May;88(5):654-660. doi: 10.1097/TA.0000000000002608.

Abstract

Background: Modeling approaches offer a novel way to detect and predict coagulopathy in trauma patients. A dynamic model, built and tested on thromboelastogram (TEG) data, was used to generate a virtual library of over 160,000 simulated RapidTEGs. The patient-specific parameters are the initial platelet count, platelet activation rate, thrombus growth rate, and lysis rate (P(0), k1, k2, and k3, respectively).

Methods: Patient data from both STAAMP (n = 182 patients) and PAMPer (n = 111 patients) clinical trials were collected. A total of 873 RapidTEGs were analyzed. One hundred sixteen TEGs indicated maximum amplitude (MA) below normal and 466 TEGs indicated lysis percent above normal. Each patient's TEG response was compared against the virtual library of TEGs to determine library trajectories having the least sum-of-squared error versus the patient TEG up to each specified evaluation time ∈ (3, 4, 5, 7.5, 10, 15, 20 minutes). Using 10 nearest-neighbor trajectories, a logistic regression was performed to predict if the patient TEG indicated MA below normal (<50 mm), lysis percent 30 minutes after MA (LY30) greater than 3%, and/or blood transfusion need using the parameters from the dynamic model.

Results: The algorithm predicts abnormal MA values using the initial 3 minutes of RapidTEG data with a median area under the curve of 0.95, and improves with more data to 0.98 by 10 minutes. Prediction of future platelet and packed red blood cell transfusion based on parameters at 4 and 5 minutes, respectively, provides equivalent predictions to the traditional TEG parameters in significantly less time. Dynamic model parameters could not predict abnormal LY30 or future fresh-frozen plasma transfusion.

Conclusion: This analysis could be incorporated into TEG software and workflow to quickly estimate if the MA would be below or above threshold value within the initial minutes following a TEG, along with an estimate of what blood products to have on hand.

Level of evidence: Therapeutic/Care Management: Level IV.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, U.S. Gov't, Non-P.H.S.

MeSH terms

  • Adult
  • Algorithms
  • Blood Coagulation Disorders / blood
  • Blood Coagulation Disorders / diagnosis*
  • Blood Coagulation Disorders / therapy
  • Blood Component Transfusion / statistics & numerical data*
  • Clinical Trials as Topic
  • Female
  • Humans
  • Male
  • Middle Aged
  • Models, Cardiovascular*
  • Platelet Activation
  • Platelet Count
  • Point-of-Care Systems / statistics & numerical data
  • Prognosis
  • Thrombelastography / instrumentation
  • Thrombelastography / statistics & numerical data*
  • Time Factors
  • Wounds and Injuries / blood
  • Wounds and Injuries / complications*
  • Wounds and Injuries / therapy
  • Young Adult