Impact of CT Injector Technology and Contrast Media Viscosity on Vascular Enhancement: Evaluation in a Circulation Phantom

Br J Radiol. 2020 May 1;93(1109):20190868. doi: 10.1259/bjr.20190868. Epub 2020 Feb 20.

Abstract

Objective: To assess the impact of piston-based vs peristaltic injection system technology and contrast media viscosity on achievable iodine delivery rates (IDRs) and vascular enhancement in a pre-clinical study.

Methods: Four injectors were tested: MEDRAD® Centargo, MEDRAD® Stellant, CT Exprès®, and CT motion using five contrast media [iopromide (300 and 370 mgI ml-1), iodixanol 320 mgI ml-1, iohexol 350 mgI ml-1, iomeprol 400 mgI ml-1]. Three experiments were performed evaluating achievable IDR and corresponding enhancement in a circulation phantom.

Results: Experiment I: Centargo provided the highest achievable IDRs with all tested contrast media (p < 0.05). Iopromide 370 yielded the highest IDR with an 18G catheter (3.15 gI/s); iopromide 300 yielded the highest IDR with 20G (2.70 gI/s) and 22G (1.65 gI/s) catheters (p < 0.05).Experiment II: with higher achievable IDRs, piston-based injectors provided significantly higher peak vascular enhancement (up to 48% increase) than the peristaltic injectors with programmed IDRs from 1.8 to 2.4 gI/s (p < 0.05).Experiment III: with programmed IDRs (e.g. 1.5 gI/s) achievable by all injection systems, Centargo, with sharper measured bolus shape, provided significant increases in enhancement of 34-73 HU in the pulmonary artery with iopromide 370 (p < 0.05).

Conclusion: The tested piston-based injection systems combined with low viscosity contrast media provide higher achievable IDRs and higher peak vascular enhancement than the tested peristaltic-based injectors. With equivalent IDRs, Centargo provides higher peak vascular enhancement due to improved bolus shape.

Advances in knowledge: This paper introduces a new parameter to compare expected performance among contrast media: the concentration/viscosity ratio. Additionally, it demonstrates previously unexplored impacts of bolus shape on vascular enhancement.

Publication types

  • Comparative Study
  • Evaluation Study

MeSH terms

  • Contrast Media / administration & dosage*
  • Contrast Media / pharmacokinetics
  • Equipment Design
  • Humans
  • Injections / instrumentation
  • Iohexol / administration & dosage
  • Iohexol / analogs & derivatives
  • Iohexol / pharmacokinetics
  • Iopamidol / administration & dosage
  • Iopamidol / analogs & derivatives
  • Iopamidol / pharmacokinetics
  • Phantoms, Imaging
  • Tomography, X-Ray Computed
  • Triiodobenzoic Acids / administration & dosage
  • Triiodobenzoic Acids / pharmacokinetics
  • Viscosity

Substances

  • Contrast Media
  • Triiodobenzoic Acids
  • iomeprol
  • Iohexol
  • iopromide
  • iodixanol
  • Iopamidol