4D Flow MRI hemodynamic benchmarking of surgical bioprosthetic valves

Magn Reson Imaging. 2020 May:68:18-29. doi: 10.1016/j.mri.2020.01.006. Epub 2020 Jan 22.

Abstract

Purpose: We exploited 4-dimensional flow magnetic resonance imaging (4D Flow), combined with a standardized in vitro setting, to establish a comprehensive benchmark for the systematic hemodynamic comparison of surgical aortic bioprosthetic valves (BPVs).

Materials and methods: 4D Flow analysis was performed on two small sizes of three commercialized pericardial BPVs (Trifecta™ GT, Carpentier-Edwards PERIMOUNT Magna and Crown PRT®). Each BPV was tested over a clinically pertinent range of continuous flow rates within an in vitro MRI-compatible system, equipped with pressure transducers. In-house 4D Flow post-processing of the post-valvular velocity field included the quantification of BPV effective orifice area (EOA), transvalvular pressure gradients (TPG), kinetic energy and viscous energy dissipation.

Results: The 4D Flow technique effectively captured the 3-dimensional flow pattern of each device. Trifecta exhibited the lowest range of velocity and kinetic energy, maximized EOA (p < 0.0001) and minimized TPGs (p ≤ 0.015) if compared with Magna and Crown, these reporting minor EOA difference s (p ≥ 0.042) and similar TPGs (p ≥ 0.25). 4D Flow TPGs estimations strongly correlated against ground-truth data from pressure transducers; viscous energy dissipation proved to be inversely proportional to the fluid jet penetration.

Conclusion: The proposed 4D Flow analysis pinpointed consistent hemodynamic differences among BPVs, highlighting the not negligible effect of device size on the fluidynamic outcomes. The efficacy of non-invasive 4D Flow MRI protocol could shed light on how standardize the comparison among devices in relation to their actual hemodynamic performances and improve current criteria for their selection.

Keywords: 4D Flow MRI; Bioprosthetic valves; Hemodynamics; In vitro setting.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aortic Valve / diagnostic imaging*
  • Aortic Valve / surgery
  • Aortic Valve Stenosis / diagnostic imaging*
  • Aortic Valve Stenosis / surgery
  • Benchmarking*
  • Bioprosthesis / standards*
  • Heart Valve Prosthesis / standards*
  • Hemodynamics
  • Humans
  • Imaging, Three-Dimensional
  • In Vitro Techniques
  • Kinetics
  • Magnetic Resonance Imaging*
  • Prosthesis Design / standards*