Purpose: Advanced 3D dosimetry is required for verifications of complex dose distributions in modern radiotherapy. Two 3D polymer gel dosimeters, coupled with magnetic resonance (MR) imaging (3 T MRI) readout and data processing with polyGeVero® software, were tested for the verification of calculated 3D dose distributions by a treatment planning system (TPS) and ArcCHECK®-3DVH®, related to eradication of a lung tumour.
Methods: N-vinylpyrrolidone-containing 3D polymer gel dosimeters were used: VIC (containing ascorbic acid and copper sulfate pentahydrate) and VIC-T (containing tetrakis(hydroxymethyl)phosphonium chloride). Three remote centers were involved in the dosimeters preparation and irradiation (Poland), and MRI (Austria). Cross beam calibration of the dosimeters and verification of a 3D dose distribution calculated with an Eclipse External Beam TPS and ArcCHECK®-3DVH® were performed. The 3D-to-3D comparisons of the VIC and VIC-T with TPS and ArcCHECK®-3DVH® along with ArcCHECK®-3DVH® versus TPS dose matrixes were performed with the aid of the polyGeVero® by analyzing dose profiles, isodoses lines, gamma index, gamma angle, dose difference, and related histograms.
Results: The measured MR-relaxation rate (R2 = 1/T2) for the dosimeters relates to the dose, as follows: R2 = 0.0928 ± 0.0008 [Gy-1 s-1] × D [Gy] + 2.985 ± 0.012 [s-1] (VIC) and 0.1839 ± 0.0044 [Gy-1 s-1] × D [Gy] + 2.519 ± 0.053 [s-1] (VIC-T). The 3D-to-3D comparisons revealed a good agreement between the measured and calculated 3D dose distributions.
Conclusions: VIC and VIC-T with 3T MRI readout and polyGeVero® showed potential for verifications of calculated irradiation plans. The results obtained suggest the implementation of the irradiation plan for eradication of the lung tumour.
Keywords: 3D radiotherapy dosimetry; 3T MRI; ArcCHECK–3DVH verification; TPS verification; VIC; VIC-T; polyGeVero software.