Quantification of fiducial marker visibility for MRI-only prostate radiotherapy simulation

Phys Med Biol. 2020 Feb 5;65(3):035015. doi: 10.1088/1361-6560/ab65db.

Abstract

To objectively compare the suitability of MRI pulse sequences and commercially available fiducial markers (FMs) for MRI-only prostate radiotherapy simulation. Most FMs appear as small signal voids in MRI images making them difficult to differentiate from tissue heterogeneities such as calcifications. In this study we use quantitative metrics to objectively evaluate the visibility of FMs in 27 patients and an anthropomorphic phantom with a variety of standard clinical MRI pulse sequences and commercially available FMs. FM visibility was quantified using the local contrast-to-noise-ratio (lCNR), the difference between the 80th and 20th percentile iso-intensity FM volumes (V fall) and the largest iso-intensity volume that can be distinguished from background: apparent-marker-volume (AMV). A larger lCNR and AMV, and smaller V fall represents a more easily identifiable FM. The number of non-marker objects visualized by each pulse sequence was calculated using FM-derived template-matching. The FM-based target-registration-error (TRE) between each MRI and the planning-CT image was calculated. Fiducial marker visibility was rated by two medical physicists with over three years of experience examining MRI-only prostate simulation images. The rater's classification accuracy was quantified using the F 1 score, which is the harmonic mean of the rater's precision and recall. These quantitative metrics and human observer ratings were used to evaluate FM identifiability in images from nine subtypes of T 1-weighted, T 2-weighted and gradient echo (GRE) pulse sequences in a 27-patient study. A phantom study was conducted to quantify the visibility of 8 commercially available FMs. In the patient study, the largest mean lCNR and AMV and, smallest normalized V fall were produced by the 3.0 T multiple-echo GRE pulse sequence (T 1-VIBE, 2° flip angle, 1.23 ms and 2.45 ms echo-times). This pulse sequence produced no false marker detections and TREs less than 2 mm in the left-right, anterior-posterior and cranial-caudal directions, respectively. Human observers rated the 1.23 ms echo-time GRE images with the best average marker visibility score of 100% and an F 1 score of 1. In the phantom study, the Gold-Anchor GA-200X-20-B (deployed in a folded configuration) produced the largest sequence averaged lCNR and AMV measurements at 16.1 and 16.7 mm3, respectively. Using quantitative visibility and distinguishability metrics and human observer ratings, the patient study demonstrated that multiple-echo GRE images produced the best gold FM visibility and distinguishability. The phantom study demonstrated that markers manufactured from platinum or iron-doped gold quantitatively produced superior visibility compared to their pure gold counterparts.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Computer Simulation
  • Fiducial Markers*
  • Gold / chemistry
  • Humans
  • Iron / chemistry
  • Magnetic Resonance Imaging / methods*
  • Male
  • Phantoms, Imaging*
  • Platinum / chemistry
  • Prostatic Neoplasms / pathology*
  • Prostatic Neoplasms / radiotherapy*
  • Radiotherapy Planning, Computer-Assisted / methods*
  • Radiotherapy, Image-Guided / methods*

Substances

  • Platinum
  • Gold
  • Iron