Using T1 as a direct detection dimension in two-dimensional time-domain NMR experiments using CWFP regime

J Magn Reson. 2020 Feb:311:106666. doi: 10.1016/j.jmr.2019.106666. Epub 2019 Dec 9.

Abstract

The transverse relaxation time (T2), measured with Carr-Purcell-Meiboom-Gill (CPMG) sequence, has been widely used to obtain the direct dimension data in two-dimension time domain NMR (2D TD-NMR). In this paper we are demonstrating that Continuous Wave Free Precession sequence, with low flip angle (CWFP-T1), can be an alternative to CPMG as direct detection dimension. CWFP-T1 is a fast single shot sequence, like CPMG, and yields an exponential signal governed predominantly by the longitudinal (T1) relaxation time. To obtain the correlations between T1 and T2 (T1-T2 maps) we are proposing the use of CPMG-CWFP-T1 pulse sequence. In this sequence CPMG encodes T2 information (indirect dimension) that modulates the CWFP-T1 (direct dimension) signal amplitudes. CPMG-CWFP-T1 experiments were compared with classical 2D sequences such as Saturation-Recovery-CPMG (SR-CPMG) and Inversion-Recovery-CPMG (IR-CPMG) sequence and yields similar results in phantom sample. The experimental time for the 2D sequences, using single scan, shows that SR-CPMG ≤ CPMG-CWFP-T1 < IR-CPMG. Experimental and simulated results demonstrated that 2D-CPMG-CWFP-T1 maps have higher resolution in T1 dimension than the techniques that uses CPMG as direct dimension. CPMG-CWFP-T1 sequence was also applied to study beef samples, and 2D maps showed higher resolution in the two fat signals than the classical IR-CPMG method.

Keywords: 2D TD-NMR pulse sequence; 2D-Laplace transform; CPMG-CWFP-T(1); T(1)-T(2) maps.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adipose Tissue / chemistry
  • Algorithms
  • Animals
  • Cattle
  • Computer Simulation
  • Electronic Data Processing
  • Food Analysis / methods*
  • Magnetic Resonance Spectroscopy / methods*
  • Meat / analysis
  • Molecular Conformation
  • Signal-To-Noise Ratio