Performance of primary diagnostic monitors (PDMs) over time

J Appl Clin Med Phys. 2019 Dec;20(12):180-185. doi: 10.1002/acm2.12759.

Abstract

In this work, we evaluated the change of primary monitor characteristics in two consecutive years. Sixty-six primary monitors were included in the analysis. The monitors were located at radiology physicians' offices and radiology reading rooms. All primary monitors were equipped with the manufacturer's built-in photometers and connected to the BarcoMediCalQA web service for manual and automatic quality control measurements. External photometer/illuminance meter (RaySafe Solo Light) was used to measure the luminance values. Measured luminance values of the TG18LN1-18 and TG18UNL80 test patterns were used to evaluate the primary monitors performance. In a comparison of the quality assurance (QA) measurement results for the same monitors that were performed within 2 years, the luminance of 25 displays remained statistically the same (P > 0.01). The luminance of 17 displays decreased (P < 0.01) in 2017 when compared with 2016, the luminance of 24 displays increased (P < 0.01) in 2017 when compared with 2016. For the annual measurements of the MLD in 2016 and 2017, 25 out of 66 displays showed a decrease of MLD values in 2017 compared with the same measurements in 2016 and 41 displays showed an increase of MLD in 2017. All tested primary displays had the MLD value less than 17.2%. The mean value of illuminance measured in 2016 was 5.8 lux ± 3.1 lux. In 2017, the mean value of illuminance measured was 8.7 lux ± 5.3 lux. Although it is expected that monitors luminance values will decrease over time, we found displays with increased luminance. This is possibly due to the multiple monitor calibrations that were performed between two annual monitor QA tests. Based on the findings of this work, more efficient display QA programs with a shorter time interval than 1 year are needed.

Keywords: grayscale display function; primary diagnostic monitor; quality assurance.

MeSH terms

  • Calibration
  • Computer Graphics / standards
  • Data Display / standards*
  • Diagnostic Imaging / instrumentation*
  • Diagnostic Imaging / standards*
  • Humans
  • Luminescent Measurements
  • Photometry / standards*
  • Quality Control*
  • Radiology Information Systems / standards*
  • Time Factors