Comparison of cord blood lactate measurement by gas analyzer and portable electrochemical devices

J Perinat Med. 2020 Feb 25;48(2):157-161. doi: 10.1515/jpm-2019-0357.

Abstract

Objective To compare the accuracy of cord blood lactate measurement using gas analyzer and portable devices in order to assess possibility of implementation of these devices in clinical practice. Methods We performed a prospective observational study using 30 umbilical cord samples which were obtained immediately after birth. Portable electrochemical devices Lactate Scout (SensLab GmbH, Leipzig, Germany) and StatStrip Lactate (NOVA Biomedical, Waltham, MA, USA) were used to determine lactate level. A gas analyzer ABL800 FLEX (Radiometer Medical ApS, Brønshøj-Husum, Denmark) was used as a reference. Base excess (BE), pH, partial oxygen (pO2) and carbon dioxide (pCO2) pressure, hemoglobin (ctHb) and bilirubin (ctBl) levels were measured. Results The mean umbilical cord blood lactate level determined by the gas analyzer was 5.85 ± 2.66 mmol/L (ranging from 1.4 mmol/L to 13.4 mmol/L). Lactate level estimated by Lactate Scout was 5.66 ± 2.65 mmol/L and did not significantly differ from the reference method level (P = 0.2547). The mean lactate level determined by StatStrip Lactate was significantly lower than by the gas analyzer - 4.81 ± 2.38 mmol/L (P < 0.0001). Umbilical cord blood pH, BE, pO2 and pCO2, ctHb and ctBl levels did not affect the accuracy of the lactate measurement in absolute units (mmol/L). Conclusion Umbilical cord blood lactate level measured by StatStrip Lactate was lower than estimated by the ABL800 FLEX gas analyzer. This shows the necessity to develop decision-making reference points separately for each device. Umbilical cord blood pH, BE, pO2 and pCO2, ctHb and ctBl levels did not affect the accuracy of measurements by electrochemical portable devices.

Keywords: acid base status; base excess; cord blood lactate; fetal hypoxia; gas analyzer; labor; pH; partial carbon dioxide pressure; partial oxygen pressure.

Publication types

  • Comparative Study
  • Observational Study

MeSH terms

  • Blood Gas Analysis / instrumentation*
  • Electrochemical Techniques / instrumentation
  • Fetal Blood / chemistry
  • Humans
  • Lactic Acid / blood*
  • Prospective Studies

Substances

  • Lactic Acid