Beta-ray imaging system with γ-ray coincidence for multiple-tracer imaging

Med Phys. 2020 Feb;47(2):587-596. doi: 10.1002/mp.13947. Epub 2019 Dec 25.

Abstract

Purpose: Beta-ray imaging systems are widely used for various biological objects to obtain a two-dimensional (2D) distribution of β-ray emitting radioisotopes. However, a conventional β-ray imaging system is unsuitable for multiple-tracer imaging, because the continuous energy distribution of β-rays complicates distinguishing among different tracers by energy information. Therefore, we developed a new type of β-ray imaging system, which is useful for multiple tracers by detecting coincidence γ-rays with β-rays, and evaluated its imaging performance.

Methods: Our system is composed of position-sensitive β-ray and γ-ray detectors. The former is a 35 × 35 × 1-mm3 Ce-Doped((La, Gd)2 Si2 O7 ) (La-GPS) scintillation detector, which has a 300-µm pitch of pixels. The latter is a 43 × 43 × 16-mm3 bismuth germanium oxide (BGO) scintillation detector. Both detectors are mounted on a flexible frame and placed in a user-selectable position. We experimentally evaluated the performance of the β-ray detector and the γ-ray efficiencies of the γ-ray detector with different energies, positions, and distances. We also conducted point sources and phantom measurements with dual isotopes to evaluate the system performance of multiple-tracer imaging.

Results: For the β-ray detector, the β-ray detection efficiencies for 45 Ca (245-keV maximum energy) and 90 Sr/90 Y (545 and 2280-keV maximum energy) were 14.3% and 21.9%, respectively. The total γ-ray detection efficiency of the γ-ray detector for all γ-rays from 22 Na (511-keV annihilation γ-rays and a 1275-keV γ-ray) in the center position with a detector distance of 20 mm was 17.5%. From a point-source measurement using 22 Na and 90 Sr/90 Y, we successfully extracted the position of a positron-γ emitter 22 Na. Furthermore, for a phantom experiment using 45 Ca and 18 F or 18 F and 22 Na, we successfully extracted the distribution of the second tracer using the annihilation γ-ray or de-excitation γ-ray coincidence. In all the imaging experiments, the event counts of the extracted images were consistent with the counts estimated by the measured γ-ray efficiencies.

Conclusions: We successfully demonstrated the feasibility of our β-ray autoradiography system for imaging multiple isotopes. Since our system can identify not only a β-γ emitter but also a positron emitter using the coincidence detection of annihilation γ-rays, it is useful for PET tracers and various new applications that are otherwise impractical.

Keywords: coincidence measurement; multiple tracers; scintillation detector; β-ray; β-ray imaging; γ-ray.

MeSH terms

  • Beta Particles
  • Bismuth / chemistry
  • Calcium Radioisotopes / chemistry*
  • Cerium / chemistry
  • Equipment Design
  • Gamma Rays
  • Germanium / chemistry
  • Image Processing, Computer-Assisted
  • Lanthanum / chemistry
  • Models, Theoretical
  • Oxygen / chemistry
  • Phantoms, Imaging*
  • Positron-Emission Tomography
  • Scintillation Counting / instrumentation*
  • Silicon / chemistry
  • Sodium / chemistry
  • Strontium Radioisotopes / chemistry*

Substances

  • Calcium Radioisotopes
  • Calcium-45
  • Strontium Radioisotopes
  • Germanium
  • bismuth germanium oxide
  • Cerium
  • Strontium-90
  • Lanthanum
  • Sodium
  • Oxygen
  • Bismuth
  • Silicon