Transcatheter mitral valve replacement (TMVR): annular or apical fixation?

EuroIntervention. 2020 Aug 7;16(6):e510-e517. doi: 10.4244/EIJ-D-19-00614.

Abstract

Aims: The aim of this study was to evaluate the impact of two different transcatheter mitral valve replacement (TMVR) fixation strategies on the neo left ventricular outflow tract (neo-LVOT) and aorto-mitral angulation (AMA) after TMVR.

Methods and results: Two different self-expanding nitinol valved stents were developed for transapical TMVR. In one group, the stents were fixed with an annular fixation system (ANN group, n=6). These prototypes were compared with an apical tether fixation TMVR system (AP group, n=11) in another group. Echocardiographic evaluation of the AMA and the neo-LVOT was conducted before and one hour after implantation. Maximal and minimal AMA (AMAmax and AMAmin) during the cardiac cycle of the AP group were significantly narrower than those of the ANN group (AMAmax: 39±8° vs 67±15°, p<0.001, AMAmin: 33±10° vs 56±22°, p=0.009). More severe reduction of the neo-LVOT diameter was observed in the ANN group (60±11% vs 26±14%, p<0.001). The ANN group had a higher peak velocity through the neo-LVOT post implantation (200±52 cm/s vs 108±15 cm/s, p<0.001).

Conclusions: The apical fixation system maintains a smaller and more stable aorto-mitral angulation and a larger neo-LVOT, thereby reducing the risk of postoperative neo-LVOT obstruction in this experimental setting.

MeSH terms

  • Cardiac Catheterization / methods*
  • Echocardiography
  • Heart Valve Prosthesis Implantation*
  • Heart Valve Prosthesis*
  • Humans
  • Mitral Valve / surgery*
  • Mitral Valve Insufficiency / diagnostic imaging
  • Mitral Valve Insufficiency / surgery*
  • Treatment Outcome
  • Ventricular Outflow Obstruction / surgery*