Diagnosis of Primary Aldosteronism by Seated Saline Suppression Test-Variability Between Immunoassay and HPLC-MS/MS

J Clin Endocrinol Metab. 2020 Mar 1;105(3):dgz150. doi: 10.1210/clinem/dgz150.

Abstract

Background: In primary aldosteronism (PA), excessive, autonomous secretion of aldosterone is not suppressed by salt loading or fludrocortisone. For seated saline suppression testing (SSST), the recommended diagnostic cutoff 4-hour plasma aldosterone concentration (PAC) measured by high-performance liquid chromatography-mass spectrometry (HPLC-MS/MS is 162 pmol/L. Most diagnostic laboratories, however, use immunoassays to measure PAC. The cutoff for SSST using immunoassay is not known. We hypothesized that the cutoff is different between the assays.

Methods: We analyzed 80 of the 87 SSST tests that were performed during our recent study defining the HPLC-MS/MS cutoff. PA was confirmed in 65 by positive fludrocortisone suppression testing (FST) and/or lateralization on adrenal venous sampling and excluded in 15 by negative FST. PAC was measured by a chemiluminescence immunoassay (PACIA) in the SSST samples using the DiaSorin Liaison XL analyzer, and receiver operating characteristics (ROC) analysis was performed to identify the PACIA cutoff.

Results: ROC revealed good performance (area under the curve = 0.893; P < .001) of 4-hour postsaline PACIA for diagnosis of PA and an optimal diagnostic cutoff of 171 pmol/L, with sensitivity and specificity of 95.4% and 80.0%, respectively. A higher cutoff of 217 pmol/L improved specificity (86.7%) with lower sensitivity (86.2%). PACIA measurements strongly correlated with PAC measured by HPLC-MS (r = 0.94, P < .001).

Conclusions: A higher diagnostic cutoff for SSST should be employed when PAC is measured by immunoassay rather than HPLC-MS/MS. The results suggest that (i) PA can be excluded if 4-hour PACIA is less than 171 pmol/L, and (ii) PA is highly likely if the PACIA is greater than 217 pmol/L by chemiluminescence immunoassay. A gray zone exists between the cutoffs of 171 and 217 pmol/L, likely reflecting a lower specificity of immunoassay.

Keywords: confirmation; diagnosis; hyperaldosteronism; immunoassay; primary aldosteronism; seated saline suppression test.

Publication types

  • Observational Study
  • Research Support, Non-U.S. Gov't
  • Validation Study

MeSH terms

  • Aldosterone / blood*
  • Aldosterone / isolation & purification
  • Chromatography, High Pressure Liquid / methods
  • Chromatography, High Pressure Liquid / standards
  • Cohort Studies
  • Female
  • Humans
  • Hyperaldosteronism / blood
  • Hyperaldosteronism / diagnosis*
  • Immunoassay / methods
  • Immunoassay / standards*
  • Luminescent Measurements / methods
  • Luminescent Measurements / standards
  • Male
  • Middle Aged
  • ROC Curve
  • Reference Values
  • Reproducibility of Results
  • Saline Solution / administration & dosage
  • Sitting Position
  • Tandem Mass Spectrometry / methods
  • Tandem Mass Spectrometry / standards*

Substances

  • Saline Solution
  • Aldosterone