Detection of specific antibody-ligand interactions with a self-induced back-action actuated nanopore electrophoresis sensor

Nanotechnology. 2019 Nov 1;31(8):085502. doi: 10.1088/1361-6528/ab53a7. Online ahead of print.

Abstract

Recent advances in plasmonic nanopore technologies have enabled the use of concurrently acquired bimodal optical-electrical data for improved quantification of molecular interactions. This work presents the use of a new plasmonic nanosensor employing self-induced back-action (SIBA) for optical trapping to enable SIBA-actuated nanopore electrophoresis (SANE) for quantifying antibody-ligand interactions. T-cell receptor-like antibodies (TCRmAbs) engineered to target peptide-presenting major histocompatibility complex (pMHC) ligands, representing a model of target ligands presented on the surface of cancer cells, were used to test the SANE sensor's ability to identify specific antibody-ligand binding. Cancer-irrelevant TCRmAbs targeting the same pMHCs were also tested as a control. It was found that the sensor could provide bimodal molecular signatures that could differentiate between antibody, ligand and the complexes that they formed, as well as distinguish between specific and non-specific interactions. Furthermore, the results suggested an interesting phenomenon of increased antibody-ligand complex bound fraction detected by the SANE sensor compared to that expected for corresponding bulk solution concentrations. A possible physical mechanism and potential advantages for the sensor's ability to augment complex formation near its active sensing volume at concentrations lower than the free solution equilibrium binding constant (K D ) are discussed.