Improved visual acuity using a retinal implant and an optimized stimulation strategy

J Neural Eng. 2019 Dec 23;17(1):016018. doi: 10.1088/1741-2552/ab5299.

Abstract

Objective: Retinal prosthetic devices hold great promise for the treatment of retinal degenerative diseases such as retinitis pigmentosa and age-related macular degeneration. Through electrical stimulation of the surviving retinal neurons, these devices evoke visual signals that are then relayed to the brain. Currently, the visual prostheses used in clinical trials have few electrodes, thus limiting visual acuity. Electrode arrays with high electrode densities have been developed using novel technologies, including diamond growth and laser machining, and these may provide a more promising route to achieve high visual acuity in blind patients.

Approach: Here, we studied the potential spatial resolution of electrical stimulation using diamond electrodes. We did this by labeling retinal ganglion cells in whole mount retina with a calcium indicator in wild-type rats and those with retinal degeneration. We imaged the ganglion cell responses to a range of stimulation parameters, including pulse duration and return electrode configuration.

Main results: With sub-retinal stimulation, in which electrodes were in contact with the intact or degenerated photoreceptor layer, we found that biphasic pulses of 0.1 ms phase duration and a local return configuration was the most effective in confining the retinal ganglion cell activation patterns, while also remaining within the safety limits of the materials and providing the best power efficiency.

Significance: These results provide an optimized stimulation strategy for retinal implants, which if implemented in a retinal prosthetic is expected to improve the achievable visual acuity.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Blindness / diagnostic imaging
  • Blindness / therapy
  • Electric Stimulation / methods
  • Electrodes, Implanted*
  • Female
  • Male
  • Microelectrodes
  • Molecular Imaging / methods
  • Rats
  • Rats, Long-Evans
  • Retina / diagnostic imaging*
  • Retina / physiology*
  • Visual Acuity / physiology*
  • Visual Prosthesis* / standards