Trends in Estimated Thyroid, Salivary Gland, Brain, and Eye Lens Doses From Intraoral Dental Radiography Over Seven Decades (1940 TO 2009)

Health Phys. 2020 Feb;118(2):136-148. doi: 10.1097/HP.0000000000001138.

Abstract

The purpose of this study is to support retrospective dose estimation for epidemiological studies by providing estimates of historical absorbed organ doses to the brain, lens of the eye, salivary glands, and thyroid from intraoral dental radiographic examinations performed from 1940 to 2009. We simulated organ doses to an adult over 10 y time periods from 1940 to 2009, based on commonly used sets of x-ray machine settings collected from the literature. Simulations to estimate organ dose were performed using personal computer x-ray Monte Carlo software. Overall, organ doses were less than 1 mGy for a single intraoral radiograph for all decades. From 1940 to 2009, doses to the brain, eye lens, salivary glands, and thyroid decreased by 86, 96, 95, and 89%, respectively. Of these four organs, the salivary glands received the highest doses, with values decreasing from about 0.23 mGy in the 1940s to 0.025 mGy in the 2000s for a single intraoral radiograph. Based on simulations using collected historical data on x-ray technical parameters, improvements in technology and optimization of the technical settings used to perform intraoral dental radiography have resulted in a decrease in absorbed dose to the brain, eye lens, salivary glands, and thyroid over the period from 1940 to 2009.

The purpose of this study is to support retrospective dose estimation for epidemiological studies by providing estimates of historical absorbed organ doses to the brain, lens of the eye, salivary glands, and thyroid from intraoral dental radiographic examinations performed from 1940 to 2009. We simulated organ doses to an adult over 10 y time periods from 1940 to 2009, based on commonly used sets of x-ray machine settings collected from the literature. Simulations to estimate organ dose were performed using personal computer x-ray Monte Carlo software. Overall, organ doses were less than 1 mGy for a single intraoral radiograph for all decades. From 1940 to 2009, doses to the brain, eye lens, salivary glands, and thyroid decreased by 86, 96, 95, and 89%, respectively. Of these four organs, the salivary glands received the highest doses, with values decreasing from about 0.23 mGy in the 1940s to 0.025 mGy in the 2000s for a single intraoral radiograph. Based on simulations using collected historical data on x-ray technical parameters, improvements in technology and optimization of the technical settings used to perform intraoral dental radiography have resulted in a decrease in absorbed dose to the brain, eye lens, salivary glands, and thyroid over the period from 1940 to 2009.

Publication types

  • Research Support, N.I.H., Intramural
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Brain / radiation effects*
  • Humans
  • Lens, Crystalline / radiation effects*
  • Monte Carlo Method
  • Radiation Dosage
  • Radiography, Dental*
  • Retrospective Studies
  • Salivary Glands / radiation effects*
  • Thyroid Gland / radiation effects*