New life for old wires: electrochemical sensor method for neural implants

J Neural Eng. 2019 Dec 11;17(1):016007. doi: 10.1088/1741-2552/ab4c69.

Abstract

Objective: Electrochemical microsensors based on noble metals can give essential information on their microenvironment with high spatio-temporal resolution. However, most advanced chemo- and biosensors lack the long-term stability for physiological monitoring of brain tissue beyond an acute application. Noble metal electrodes are widely used as neural interfaces, particularly for stimulating in the central nervous system. Our goal was to recruit already deployed, unmodified noble metal electrodes (Pt, Pt/Ir) as in situ chemical sensors.

Approach: With advanced electrochemical sensor methods, we investigated electrode surface processes, oxidizable species and oxygen as an indicator for tissue mass transport. We developed a unique, multi-step, amperometric/potentiometric sensing procedure derived from the investigation of Pt surface processes by chronocoulometry providing fundamental characterization of the electrode itself.

Main results: The resulting electrochemical protocol preconditions the electrode, measures oxidizable and reducible species, and the open circuit potential (OCP). A linear, stable sensor performance was demonstrated, also in the presence of proteins, validating signal stability of our cyclic protocol in complex environments. We investigated our sensor protocol with microelectrodes on custom Pt/Ir-wire tetrodes by in vivo measurements in the rat brain for up to four weeks. Results showed that catalytic activity of the electrode is lost over time, but our protocol is repeatedly able to both quantify and restore electrode sensitivity in vivo.

Significance: Our approach is highly relevant because it can be applied to any existing Pt electrode. Current methods to assess the brain/electrode microenvironment mainly rely on imaging techniques, histology and analysis of explanted devices, which are often end-point methods. Our procedure delivers online and time-transient information on the chemical microenvironment directly at the electrode/tissue interface of neural implants, gives new insight into the charge transfer processes, and delivers information on the state of the electrode itself addressing long-term electrode degradation.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Biosensing Techniques / instrumentation*
  • Biosensing Techniques / methods*
  • Brain / physiology
  • Brain / surgery
  • Electrochemical Techniques / instrumentation*
  • Electrochemical Techniques / methods*
  • Electrodes
  • Female
  • Implantable Neurostimulators*
  • Microelectrodes
  • Platinum*
  • Potentiometry / instrumentation
  • Potentiometry / methods
  • Rats
  • Rats, Wistar
  • Stereotaxic Techniques / instrumentation

Substances

  • Platinum