Technical Note: Increased photon starvation artifacts at low helical pitch in ultra-low-dose CT

Med Phys. 2019 Dec;46(12):5538-5543. doi: 10.1002/mp.13845. Epub 2019 Oct 21.

Abstract

Purpose: The aim of this study was to demonstrate that a low helical pitch causes increased photon starvation artifacts at ultra-low-dose CT.

Methods: A cylindrical water phantom with a diameter of 30 cm was scanned on two different generation CT scanners: a 64-slice scanner (Sensation 64, Siemens Healthcare) and a 192-slice scanner (Somatom Force, Siemens Healthcare) at multiple effective mAs levels (mAs/pitch = 200, 100, 50, 25, and 12). The corresponding CTDIvol values were 4.1, 2.0, 1.0, 0.5 mGy, on the 64-slice scanner and 3.8, 1.9, 1.0, 0.5 mGy on the 192-slice scanner, for the selected effective mAs values. For each dose setting, the scan was repeated at four helical pitches: 1.2, 0.9, 0.6, and the lowest achievable pitch on each scanner. The tube current was automatically adjusted by the scanner so that the effective mAs, and thus CTDIvol , were kept the same for different pitches. All CT data sets were reconstructed with a slice thickness of 3mm and a medium smooth kernel. Images acquired at the same dose level but different helical pitches were visually inspected to assess photon starvation artifacts and noise levels.

Results: At the same radiation dose, image noise increased with the decreasing helical pitch. The increase was more severe on the old-generation 64-slice scanner. Photon starvation artifacts were evident at 200 effective mAs on the 64-slice scanner at 80 kV. On the 192-slice scanner there was no visible photon starvation artifacts at both 200 and 50 effective mAs (CTDIvol = 4.1 mGy and 1.0 mGy, respectively); nor was there a visible impact from the lower helical pitch. Only when the dose was lowered to be extremely low (~0.26 mGy, achievable at 70 kV), did photon starvation artifacts become evident.

Conclusions: A low helical pitch may increase image noise and photon starvation artifacts compared to a higher pitch for the same dose level, particularly at ultra-low dose CT.

Keywords: CT protocol optimization; artifacts; image quality; ultra-low dose CT.

MeSH terms

  • Artifacts*
  • Phantoms, Imaging
  • Photons*
  • Radiation Dosage*
  • Tomography, X-Ray Computed*